FDA Adverse Event Malfunction Summary report: N

*

MDR report key: 176000 · Received July 8, 1998

Report

Report Number
MW1014150
Event Type
Malfunction
Date Received
July 8, 1998
Date of Event
May 29, 1998
Report Date
July 7, 1998
Manufacturer
MARQUEST MEDICAL PRODUCTS, INC.
Product Code
BZE
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

THIS IS A FOLLOW-UP REPORT TO MEDWATCH VOLUNTARY REPORT WITH ACCESS NUMBER 1014150, SUBMITTED ON 07/07/98: STAFF MEMBER DISCOVERED A SMALL HOLE IN THE EXPIRATORY LIMB OF THE PT CIRCUIT. THE APPROX 2" OF DANGLING WIRE (BETWEEN THE CIRCUIT AND THE HEATER BUT NOT WITHIN THE TUBING), HAD BEEN DRAPPED OVER THE EXPIRATORY LIMB, AND THE CIRCUIT APPEARS TO HAVE MELTED IN THAT VICINITY, FROM THE OUTSIDE INWARD. A STAFF MEMBER SPOKE WITH A PRODUCT MGR FROM THE MFR, WHO EXPLAINED THAT WHEN THE PT IS DISCONNECTED FROM THE CIRCUIT, THE LACK OF AIRFLOW THROUGH THE EXPIRATORY LIMB CAN RESULT IN THE WIRE HEATING UP. SHE ACKNOWLEDGED HAVING SEEN THIS PROBLEM WITH THIS PRODUCT SEVERAL TIMES IN THE PAST FEW YRS. NO NEGATIVE PT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 * INFANT HEATED WIRE CIRCUIT (VENTILATOR) BZE MARQUEST MEDICAL PRODUCTS, INC. PART # 155518 *

Patients

Seq Age Sex Outcome Treatment
1 *