MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2023-09835
- Event Type
- Injury
- Date Received
- August 22, 2023
- Date of Event
- May 10, 2023
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317000297
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON MARCH 29, 2024, MENTOR BECAME AWARE THAT THE PATIENT ALSO EXPERIENCED BILATERAL PTOSIS IN ADDITION TO RIGHT SIDE CAPSULAR CONTRACTURE; BAKER GRADE III, AND UNDERWENT RIGHT SIDE REPLACEMENT SURGERY WITH CATALOG NUMBER 3503251BC; SERIAL NUMBER (B)(6) ON (B)(6) 2024. ON LEFT SIDE MASTOPEXY WAS PERFORMED. THIS SUPPLEMENTAL MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT-SIDED DEVICE. LEFT SIDE COMPLICATION IS BEING REPORTED SEPARATELY. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON APRIL 3, 2024, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON APRIL 10, 2024, DEVICE EVALUATION WAS COMPLETED AS FOLLOWS: MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. DURING VISUAL EVALUATION NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED ON THE SM MPP GEL 325CC RETURNED DEVICE. MENTOR CONCLUDED THAT THTTPS://EMDR.FDA.GOV/EMDR/FORMREDACTION/B5.JSF#HE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. SINCE NO MALFUNCTION WAS OBSERVED DURING THE INVESTIGATION, NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. THIS SUPPLEMENTAL MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT-SIDED DEVICE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: AS OF NOW THERE IS NO INFORMATION REGARDING THE EXPLANT OR REOPERATION. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON (B)(6) 2023, MENTOR BECAME AWARE THAT THE SURGERY IS SCHEDULED FOR MARCH 8, 2024. HENCE, FOLLOWING FIELDS HAVE BEEN UPDATED ON THIS FORM: IS REQUIRED INTERVENTION HAS BEEN SELECTED UNDER SECTION B; FIELD B2 FIELDS D6B FOR EXPLANTATION DATE IS (B)(6) 2024. FIELD H6 HEALTH EFFECT - IMPACT CODE ¿DEVICE EXPLANTATION¿ HAS BEEN ADDED FIELD H6 TYPE OF INVESTIGATION HAS BEEN UPDATED TO "DEVICE NOT RETURNED" REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RIGHT CAPSULAR CONTRACTURE; BAKER GRADE III SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A 54-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT BREAST SURGERY WITH 325CC MENTOR MEMORYGEL BREAST IMPLANT AND EXPERIENCED RIGHT SIDE CAPSULAR CONTRACTURE; BAKER GRADE III POSTOPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310598 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 9634870 | 00081317000297 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Female | Required Intervention| O |