FDA Adverse Event Malfunction Summary report: N

CERTAS PLUS INLINE VALVE ONLY

MDR report key: 17596155 · Received August 22, 2023

Report

Report Number
3013886523-2023-00283
Event Type
Malfunction
Date Received
August 22, 2023
Date of Event
July 31, 2023
Report Date
October 13, 2023
Manufacturer
INTEGRA LIFESCIENCES SWITZERLAND SAR
Product Code
JXG
UDI-DI
10381780521013
PMA / PMN Number
K143111
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

UPDATED FIELDS: D4, D9, G3, G6, H2, H3, H4, H6, H10. THE CERTAS VALVE ((B)(6)) WAS RETURNED FOR EVALUATION. DEVICE HISTORY RECORD (DHR) - PRODUCT CODE 82-8800PL WITH LOT 7052647 CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK. FAILURE ANALYSIS - THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS AT SETTING 4. THE VALVE WAS VISUALLY INSPECTED, THE INNER BLISTER LID WAS SEALED WITH TAPE THE TAPE WAS CUT AND BLISTER LID REMOVED, A HAIR WAS FOUND ON THE UNDERSIDE OF THE VALVE. NOT VISIBLE FROM THE BLISTER SIDE. ROOT CAUSE ANALYSIS- THE ROOT CAUSE FOR THE ISSUE REPORTED BY THE CUSTOMER IS DUE TO HUMAN ERROR. THE MANUFACTURING DEPARTMENT WERE INFORMED OF THE ISSUE AND AN AWARENESS WAS PERFORMED.

Additional Manufacturer Narrative · 0

AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

N/A

Description of Event or Problem · 0

2 OF 2 REPORTS. OTHER MFG REPORT NUMBER: 3013886523-2023-00282. A FACILITY REPORTED A HAIR IN THE STERILE PACKAGING OF A CERTAS VALVE (ID 828800PL) AND THE BACTISEAL VENTRICULAR CATHETER (ID 823073) WAS CONTAMINATED BECAUSE OF THE ASSOCIATED CONTAMINATED VALVE. IT WAS DISCOVERED DURING SETUP FOR A VP SHUNT CASE ON (B)(6) 2023 AND THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT PRODUCT AVAILABLE. NO PATIENT CONTACT/INJURY AND THE EVENT DID NOT LED TO SURGICAL DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300789 CERTAS PLUS INLINE VALVE ONLY CERTAS PLUS JXG INTEGRA LIFESCIENCES SWITZERLAND SAR 7052647 10381780521013

Patients

Seq Age Sex Outcome Treatment
1 Unknown