FDA Adverse Event Malfunction Summary report: N

WARMTOUCH COMPACT WARMING UNIT

MDR report key: 1759603 · Received July 13, 2010

Report

Report Number
2936999-2010-01021
Event Type
Malfunction
Date Received
July 13, 2010
Date of Event
May 27, 2010
Report Date
June 16, 2010
Manufacturer
COVIDIEN / FORMERLY TYCO HEALTHCARE
Product Code
DWJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE UNIT WAS RETURNED TO THE COVIDIEN SERVICE DEPARTMENT IN THE UNITED KINGDOM FOR DIAGNOSIS AND PERFORMANCE VERIFICATION TESTING (PVT). ALL ELECTRICAL SAFETY TESTS WERE PERFORMED TO THE TEST INSTRUCTION SPECIFICATIONS. THE UNIT PERFORMED AS INTENDED AND TEMPERATURE RANGE WAS WITHIN SPECIFICATION. THE FAILURE WAS NOT CONFIRMED. THE POWER CORD WAS FOUND TO BE BROKEN OR IN POOR CONDITION AND WAS REPLACED PER SERVICING. THERE WAS NO TEST DATA PROVIDED BY THE HOSPITAL ENGINEERS TO CONFIRM ANY DIAGNOSTIC TESTING WAS PERFORMED IN SUPPORT OF THEIR CONCLUSIONS REGARDING THE CAUSE OF THE REPORTED FAILURE. NOTE: THE WARMTOUCH (B)(4) IS NOT DISTRIBUTED IN THE U.S. HOWEVER, THE WARMTOUCH 5300 IS OF ESSENTIALLY IDENTICAL DESIGN AND IS DISTRIBUTED IN THE U.S. THE 510K NUMBER FOR THE WARMTOUCH 5300 IS K020604.

Description of Event or Problem · 1

THE LOCAL HOSPITAL TECHNICAL ENGINEERS REPORTED TO COVIDIEN SERVICE DEPARTMENT ENGINEERS THE ALARM ON THE WARMTOUCH DID NOT COME ON, THE TEMPERATURE WAS ABOVE SPECIFICATION AND IT DID NOT PERFORM THE STEP DOWN TO THE LOWER TEMPERATURE, THEY REPORTED THE CAUSE AS EITHER THE MAIN PCB, THERMISTOR SENSOR, OR BOTH. THERE WAS NO INFORMATION PROVIDED FROM THE TECHNICAL ENGINEERS TO SUGGEST THE EVENT OCCURRED DURING ANY PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WARMTOUCH COMPACT WARMING UNIT CONVECTIVE AIR WARMER DWJ COVIDIEN / FORMERLY TYCO HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1