AUTOFEED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2010-00389
- Event Type
- Malfunction
- Date Received
- June 23, 2010
- Date of Event
- May 20, 2010
- Report Date
- May 27, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). METHOD: THE MR290 CHAMBER WAS VISUALLY INSPECTED FOR DAMAGE OR DEFORMITIES. RESULTS: ONE OF THE TWO PORTS ON THE MR290 CHAMBER WAS BENT INWARDS. THE AREA AROUND THE BASE OF THE DEFORMED CHAMBER PORT WAS SMOOTH AND FREE OF STRESS MARKS. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 090223. CONCLUSION: THE CHAMBER PORT WAS DEFORMED AND BENT OVER. THE AREA AROUND THE PORT WAS SMOOTH SUGGESTING THAT THE PORT HAD DEFORMED DUE TO EXPOSURE TO HEAT RATHER THAN FORCE, WHICH WOULD HAVE CAUSED STRESS MARKS OR CRACKING. IT IS LIKELY THAT A FLOW RESTRICTOR VALVE WAS ATTACHED TO THE CHAMBER PORT AS PART OF THE BREATHING CIRCUIT SET-UP. IN THE EVENT THAT FLOW THROUGH THE CHAMBER IS CEASED FOR A PERIOD OF TIME (DUE TO INCORRECT VENTILATOR SETTINGS OR FLOW RESTRICTOR VALVE FAILURE), AN INCREASE OF TEMPERATURE INSIDE OF THE CHAMBER CAN OCCUR. IN THIS EVENT, THE CHAMBER CAN SOFTEN AND BE BENT BY THE WEIGHT OF THE RESTRICTOR ON TOP OF THE PORT. THE DETAILS OF HUMIDIFIER WERE NOT SUPPLIED WITH THE COMPLAINT HOWEVER ALL FISHER AND PAYKEL HEALTHCARE HUMIDIFIERS CONTAIN BOTH SOFTWARE AND MANUAL TEMPERATURE CONTROL MEASURES INCLUDING THERMAL CUT OUT SWITCHES TO PREVENT OVER HEATING.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT ONE OF THE PORTS ON AN MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS DEFORMED. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOFEED HUMIDIFICATION CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE, LTD. | MR290V | 090223 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |