FDA Adverse Event Malfunction Summary report: N

AUTOFEED HUMIDIFICATION CHAMBER

MDR report key: 1759397 · Received June 23, 2010

Report

Report Number
9611451-2010-00389
Event Type
Malfunction
Date Received
June 23, 2010
Date of Event
May 20, 2010
Report Date
May 27, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE MR290 CHAMBER WAS VISUALLY INSPECTED FOR DAMAGE OR DEFORMITIES. RESULTS: ONE OF THE TWO PORTS ON THE MR290 CHAMBER WAS BENT INWARDS. THE AREA AROUND THE BASE OF THE DEFORMED CHAMBER PORT WAS SMOOTH AND FREE OF STRESS MARKS. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 090223. CONCLUSION: THE CHAMBER PORT WAS DEFORMED AND BENT OVER. THE AREA AROUND THE PORT WAS SMOOTH SUGGESTING THAT THE PORT HAD DEFORMED DUE TO EXPOSURE TO HEAT RATHER THAN FORCE, WHICH WOULD HAVE CAUSED STRESS MARKS OR CRACKING. IT IS LIKELY THAT A FLOW RESTRICTOR VALVE WAS ATTACHED TO THE CHAMBER PORT AS PART OF THE BREATHING CIRCUIT SET-UP. IN THE EVENT THAT FLOW THROUGH THE CHAMBER IS CEASED FOR A PERIOD OF TIME (DUE TO INCORRECT VENTILATOR SETTINGS OR FLOW RESTRICTOR VALVE FAILURE), AN INCREASE OF TEMPERATURE INSIDE OF THE CHAMBER CAN OCCUR. IN THIS EVENT, THE CHAMBER CAN SOFTEN AND BE BENT BY THE WEIGHT OF THE RESTRICTOR ON TOP OF THE PORT. THE DETAILS OF HUMIDIFIER WERE NOT SUPPLIED WITH THE COMPLAINT HOWEVER ALL FISHER AND PAYKEL HEALTHCARE HUMIDIFIERS CONTAIN BOTH SOFTWARE AND MANUAL TEMPERATURE CONTROL MEASURES INCLUDING THERMAL CUT OUT SWITCHES TO PREVENT OVER HEATING.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT ONE OF THE PORTS ON AN MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS DEFORMED. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOFEED HUMIDIFICATION CHAMBER BTT FISHER & PAYKEL HEALTHCARE, LTD. MR290V 090223

Patients

Seq Age Sex Outcome Treatment
1