BABY CONTROLLED WALL MOUNT INFANT WARMER
Report
- Report Number
- 9611451-2010-00393
- Event Type
- Malfunction
- Date Received
- June 23, 2010
- Report Date
- May 25, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- FMT
- PMA / PMN Number
- K971461
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). METHOD: THE WALL MOUNT WARMER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE'S REGIONAL OFFICE AND SERVICE CENTER IN (B)(4). THE WARMER WAS VISUALLY INSPECTED AND TESTED FOR FUNCTIONALITY. THE HARNESS AND HEATING ELEMENT COMPONENTS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE ((B)(4)) FOR FURTHER INSPECTION AND RESISTANCE TESTED USING A MULTIMETER. RESULTS: THERE WAS NO DAMAGE TO THE EXTERNAL CONTROLLER ASSEMBLY OR WARMER HEAD. UPON FURTHER INSPECTIONS AND OPENING OF THE UNIT IT WAS DISCOVERED THAT TWO OF THE INTERNAL ELECTRICAL CONNECTORS WERE DAMAGED. WHEN THE WARMER WAS POWERED ON, IT DISPLAYED AN ERROR CODE TO ALERT THE USER TO THE FAULT CONDITION AND NO HEAT WAS GENERATED. THE RESISTANCE OF THE HEATING ELEMENT WAS FOUND TO BE NORMAL. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 070615. CONCLUSION: THE HEATER COMPONENTS WERE FOUND TO BE DAMAGED. IT IS POSSIBLE THAT CORROSION OR A LOOSE TERMINAL IN THE HEATER ELEMENT CAUSED INCREASED RESISTANCE, RESULTING IN A FAULTY CONNECTION AND INCREASED HEAT GENERATION IN THE WARMER HEAD. THE WARMER DISPLAYED AN ERROR CODE AND ENTERED A FAIL-SAFE STATE WHERE THE HEATER ELEMENT IS DISABLED AND THE INFANT WARMER ALARMS TO ALLOW THE USER TO ACT AND PROVIDE AN ALTERNATE MEANS OF WARMING. ALL COMPONENTS ON THE HARNESS ASSEMBLY OF THE INFANT WARMER ARE ENCLOSED IN A SHEATH AND FIRE RATED BY (B)(4).
A HOSPITAL IN (B)(6) REPORTED THAT AN IW980 WALL MOUNT INFANT WARMER "HEATS INTERMITTENTLY". NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BABY CONTROLLED WALL MOUNT INFANT WARMER | FMT | FISHER & PAYKEL HEALTHCARE LTD | IW980JEU | 070615 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |