PATIENT INFORMATION CENTER IX
Report
- Report Number
- 1218950-2023-00588
- Event Type
- Injury
- Date Received
- August 21, 2023
- Date of Event
- July 25, 2023
- Report Date
- August 6, 2025
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- MHX
- UDI-DI
- 00884838093041
- PMA / PMN Number
- K183387
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
PHILIPS IS IN PROCESS OF OBTAINING ADDITIONAL INFORMATION. A FINAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
VISUAL INSPECTION FOUND THIS WAS A ONETIME INCIDENT PER THE FIELD SERVICE ENGINEER. THE CUSTOMER HAS NOT REPORTED ANY FURTHER INCIDENTS. BASED ON THE INFORMATION AVAILABLE AND THE TESTING CONDUCTED WE WERE UNABLE TO REPLICATE THE REPORTED PROBLEM. THE REPORTED PROBLEM WAS NOT CONFIRMED. THE DEVICE WAS CONFIRMED TO BE OPERATING PER SPECIFICATIONS AND NO FAILURE WAS IDENTIFIED. THE INVESTIGATION CONCLUDES THAT NO FURTHER ACTION IS REQUIRED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED THE COMPLAINT FILE WILL BE REOPENED.
B1: ADVERSE EVENT/ PRODUCT PROBLEM IS CHANGED FROM ADVERSE EVENT TO BOTH B2: OUTCOMES ATTRIBUTED TO AE IS CHANGED FROM DEATH TO LIFE THREATENING B3: DATE OF EVENT IS ADDED H4: MANUFACTURE DATE IS UPDATED TO REFLECT THE RELEASE OF THE SOFTWARE ADDITIONAL ANALYSIS OF THE PROVIDED INFORMATION WAS PERFORMED AND THE INVESTIGATION IS UPDATED. IT WAS FURTHER CLARIFIED THAT THE SECTORS ON THE CENTRAL STATION WERE NOT INDICATING WHAT TYPE OF ALARM WAS SOUNDING. IT WAS ALSO INDICATED THE CLINICIAN WOULD NOT RESPOND TO THEIR PATIENT'S NEEDS IF AN AUDIBLE ALARM SOUNDED WITH NO ALARM TYPE SPECIFIED. THE CUSTOMER FEEDBACK REPORTING THIS ISSUE TOGGLED YES TO ACTUAL HARM. HOWEVER SUBSEQUENT REQUESTS FOR DETAILS OF THIS HARM INDICATED THAT IT IS UNKNOWN IF THERE WAS ACTUAL PATIENT HARM. THIS RECORD WILL BE DOCUMENTED AS A SERIOUS INJURY OUT OF AN ABUNDANCE OF CAUTION. THE PIC IX WAS IN USE WITH AN MX40 TELEMETRY DEVICE, AS WELL AS AN INTELLIVUE MX450 BEDSIDE MONITOR. THE CLINICAL AUDIT LOGS WERE RETRIEVED FOR KLARMAN 9N AND WERE REVIEWED FOR BED K09330 BY A PHILIPS CLINICAL PRODUCT SPECIALIST (CPS). THE CPS IDENTIFIED THAT THIS BED HAD MULTIPLE RED AND YELLOW ALARMS AROUND THE STATED TIME FRAME OF 9:00 AM. A REVIEW OF THE TECHNICAL LOGS REVEALED THAT THE ECG SOURCE WAS SWITCHED FROM THE MX40 TELEMETRY DEVICE TO THE MX450 BEDSIDE MONITOR PER UNIT WORKFLOW/PATIENT NEEDS. WHEN THE ECG LEADS WERE REMOVED FROM THE MX40 AND CONNECTED TO THE MX450, THE MX40 DEVICE REMAINED ON. THE MX40 CONTINUED TO SOURCE THE RED !!!ECG LEADS OFF INOP TO THE PIC IX CAUSING THE PIC IX TO ANNOUNCE A RED ALARM REMINDER EVERY 3 MINUTES EVEN THOUGH THERE WAS A VALID ECG SIGNAL FROM THE MX450. THE RED ALARM REMINDERS CONTINUED UNTIL THE MX40 DEVICE WAS PLACED IN STANDBY OR THE BATTERY WAS REMOVED. GIVEN THE DELAY SINCE THE ORIGINAL REPORT, THE CPS AND PRODUCT SUPPORT ENGINEER (PSE) REVIEWED A 30-DAY CLINICAL AUDIT LOG FROM MARCH/APRIL 2024 ON KLARMAN 9N. THERE ARE A FEW INSTANCES OF CONTINUAL RED ALARM REMINDERS FROM THE MX40 WHILE THE MX40 IS NOT THE ECG SOURCE. WHEN THIS OCCURS, THE SECTOR TURNS BLUE, THE RED ALARM SOUNDS FOR ABOUT 6 SECONDS AND THEN THE SECTOR CHANGES BACK TO BLACK. THERE IS NOT A VISIBLE !!!ECG LEADS OFF INOP BANNER AS THERE IS A VALID ECG FROM THE OTHER ASSIGNED DEVICE. THE PIC IX, MX40, MX450 ARE DESIGNED TO BE ABLE TO SHOW THE ECG SIGNAL FROM WHICHEVER MONITOR IS ASSIGNED TO THE SECTOR AND ALWAYS SHOW THE VALID SOURCE. IF MULTIPLE ECG SOURCES ARE DETECTED A !!CHECK ECG SOURCE INOP IS GENERATED. PHILIPS RECOMMENDS REMOVING THE MX40 BATTERY/BATTERIES OR PUTTING THE MX40 IN STANDBY WHEN SWITCHING THE ECG SOURCE TO ANOTHER MONITORING DEVICE. WHEN REVIEWING THE LOG DATA FOR THE ENTIRE UNIT, THERE ARE OTHER BEDS THAT PUT THE MX40 INTO STANDBY OR REMOVED THE BATTERY SHORTLY AFTER SWITCHING THE ECG SOURCE FROM THE MX40 TO THE MX450. WHEN THESE ACTIONS ARE PART OF THE STANDARD CLINICAL WORKFLOW, RED ALARM REMINDERS FROM THE MX40 !!!ECG LEADS OFF DO NOT CONTINUE TO OCCUR. BASED ON THE INFORMATION AVAILABLE AND THE TESTING CONDUCTED, THE CAUSE OF THE REPORTED PROBLEM WAS DIFFERENT CLINICAL ACTIONS TAKEN WHEN SWITCHING THE ECG FROM THE MX40 TO THE MX450 LEADING TO USER MISUNDERSTANDING. THE REPORTED PROBLEM WAS NOT CONFIRMED. A RESPONSE LETTER WAS PROVIDED TO THE CUSTOMER TO ADDRESS THE ISSUE.
THE CUSTOMER REPORTED THAT UNSPECIFIED PATIENT HARM OCCURRED RELATED TO ALARM TYPE NOT SHOWING AT THE CENTRAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1280713 | PATIENT INFORMATION CENTER IX | PATIENT INFORMATION CENTER IX | MHX | PHILIPS MEDICAL SYSTEMS | PATIENT INFORMATION CENTER IX | 00884838093041 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Death| L| O |