FDA Adverse Event Injury Summary report: N

LINER 28 MM I.D.

MDR report key: 17584608 · Received August 21, 2023

Report

Report Number
0002648920-2023-00190
Event Type
Injury
Date Received
August 21, 2023
Date of Event
July 26, 2023
Report Date
November 28, 2023
Manufacturer
ZIMMER MANUFACTURING B.V.
Product Code
KWY
UDI-DI
00889024633360
PMA / PMN Number
K833991
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). UPON RECEIVING ADDITIONAL INFORMATION OF THE REPORTED EVENT, IT WAS DETERMINED TO BE NOT REPORTABLE AS THIS PRODUCT WAS NOT INVOLVED IN THE EVENT. THE INITIAL REPORT SHOULD BE VOIDED.

Additional Manufacturer Narrative · 0

(B)(4). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001822565-023-02177, 002648920-2023-00196. D10/D11: CONCOMITANT MEDICAL PRODUCTS: CAT #: 802202802 / FEMORAL HEAD 12/14 TAPER 28 MM DIAMETER +0 MM NECK LENGTH / LOT #: 3151441. G2: AUSTRALIA. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS. HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 0

UPON RECEIVING ADDITIONAL INFORMATION OF THE REPORTED EVENT, IT WAS DETERMINED TO BE NOT REPORTABLE AS THIS PRODUCT WAS NOT INVOLVED IN THE EVENT. THE INITIAL REPORT SHOULD BE VOIDED.

Description of Event or Problem · 0

IT WAS REPORTED A PATIENT UNDERWENT A RIGHT HIP REVISION APPROXIMATE EIGHT DAYS POST IMPLANTATION DUE TO A DISLOCATION AND FRACTURE TO THE NECK OF THE FEMUR.THE SURGEON NOTED THAT THE STEM HAD EXCESSIVE ANTEVERSION, THE BIPOLAR SHELL WAS TOO BIG. THE HEAD, LINER AND STEM WERE REMOVED AND REPLACED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
491576 LINER 28 MM I.D. PROSTHESIS, HIP KWY ZIMMER MANUFACTURING B.V. NI 65745198 00889024633360

Patients

Seq Age Sex Outcome Treatment
1 83 YR Male Hospitalization| R SEE H10 NARRATIVE