INSUFFLATOR TUBING SET
Report
- Report Number
- 3010437008-2023-00005
- Event Type
- Malfunction
- Date Received
- August 21, 2023
- Date of Event
- January 8, 2019
- Report Date
- June 30, 2023
- Manufacturer
- ADVANCED MEDICAL DESIGN CO., LTD.
- Product Code
- NKC
- UDI-DI
- 04711605992114
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- 501
Narratives
IN GENERAL, THE AIR FLOW RATE OF INSUFFLATION TUBING SET WOULD BE 20 L/MIN DURING THE CLINICAL SITUATION. THEREFORE, WE SET A PRESSURE OF AIR OUTPUT AS 20 L/MIN AND THEN APPLY TO RETURNED SAMPLE TO SEE WHETHER OCCLUSION HAPPENS. THE INTENDED USE OF INSUFFLATION TUBING SET IS TO FILTER AND DELIVER INSUFFLATION GAS (CO2) FROM AN INSUFFLATION PUMP TO AN INSUFFLATION NEEDLE OR SUITABLE LAPAROSCOPIC PORT AS A MEANS TO PROVIDE PERITONEAL INSUFFLATION AND ALLOW FOR THE MANIPULATION OF INSTRUMENTS DURING LAPAROSCOPIC PROCEDURES. AND THE DESCRIPTION OF THE COMPLAINT MENTIONED THAT THE DEVICE WAS OCCLUDED. HOWEVER, AFTER INVESTIGATING THE RETURNED SAMPLE, IT WAS QUIET GOOD TO PROVIDE THE FUNCTION OF INTENDED USE AND REACH THE RATE OF CLINICAL REQUIREMENT. THE MALFUNCTION WAS NOT FOUND, BESIDES, THERE WAS NO FURTHER INFORMATION OR SCENARIOS DURING THE USE GAVE TO AMD. SUMMARY ACCORDING TO INVESTIGATION ABOVE, THE FUNCTION OF INSUFFLATION TUBING SET IS GOOD AND PROVIDE WHAT IT SHOULD DO FOR CLINICAL REQUIREMENT.
DESCRIBE THE EVENT OR PROBLEM: DURING USE THE INSUFFLATION TUBING OCCIUDED. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? : LAP APPY. WHAT PROBLEM DLD THE USER HAVE? DEVICE MALFUNCTION THAT IS, THE DEVICE DID NOT DO WHAT IT WAS SUPPOSED TO DO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1982589 | INSUFFLATOR TUBING SET | INSUFFLATOR TUBING | NKC | ADVANCED MEDICAL DESIGN CO., LTD. | TM6000 | 1927955 | 04711605992114 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | Male |