TABLET PC TREX HD MONITORING - REPAIRED
Report
- Report Number
- 9612330-2023-00029
- Event Type
- Malfunction
- Date Received
- August 18, 2023
- Date of Event
- July 24, 2023
- Report Date
- September 13, 2023
- Manufacturer
- NATUS MEDICAL INCORPORATED
- Product Code
- GWQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
INITIAL REPORT REF NATUS COMPLAINT# (B)(4). THERE ARE NO CAPA'S RELATED TO THIS ISSUE. THIS COMPLAINT DOES NOT IDENTIFY A DEFICIENCY IN THE PRODUCT DESIGN AND THEREFORE A CAPA IS NOT REQUIRED. PER DOC-010378 HAZARD ID 6.3, SEVERITY 11-NEGLIGIBLE RISK LEVEL MEDIUM. A RISK REVIEW IS NOT REQUIRED AS THIS COMPLAINT DOES NOT DESCRIBE A NEW FAILURE MODE OR NEW HARM AND THE EXISTING HAZARD SEVERITY AND/OR PROBABILITY OF OCCURRENCE HAS NOT CHANGED. PER QMS-004442, COMPLAINT HISTORIES ARE REVIEWED ROUTINELY PER QUALITY SYSTEM REQUIREMENTS AND ANY COMPLAINT TRENDS ARE ASSESSED AND DOCUMENTED AS PART OF THESE REVIEWS. DEVICE HISTORY RECORD WAS REVIEWED, AND NO RELATED FAULTS/ISSUE FOUND. UDI NUMBER: NOT APPLICABLE. INSTALL DATE: 11 NOVEMBER 2019. (B)(6) 2023: ALLEGEDLY DEFECTIVE TABLET WAS RETURNED FOR EVALUATION. AWAITING INVESTIGATION RESULTS.
INITIAL REPORT REF NATUS COMPLAINT# (B)(4). THE DEFECTIVE PART WAS RETURNED. ENGINEERING LOOKED AT THE ALLEGEDLY DEFECTIVE TABLET AND NOTED THE FOLLOWING: TABLET IS NOT POWERING UP. ROOT CAUSE WAS DETERMINED TO BE "TABLETS DON'T HAVE ANY PHYSICAL DAMAGE, IT LOOKS LIKE BAD BATCH OF TABLES FROM SUPPLIER / MANUFACTURER. " ENGINEERING RECOMMENDED THE TABLETS BE RETURNED TO THE SUPPLIER. RECOMMENDED ACTIONS: RETURN TO SUPPLIER IF TABLETS ARE UNDER WARRANTY. CLOSURE RATIONALE: COMPLAINT VERIFIED, BEING TRACKED AS A TREND. COMPLAINT WILL BE INCLUDED IN TRENDING DATA FOR FURTHER REVIEW.
PART 019761R TABLET PC TREX HD MONITORING - REPAIRED. THE CUSTOMER STATES THAT THE 'TABLET IS HEATING UP TO A DANGEROUS LEVEL'. CUSTOMER MENTIONED MULTIPLE TABLETS 4 IN TOTAL. THE TABLETS THEY HAVE ARE HEATING UP AND KEEP ON RESTARTING, CAUSING THE LOST OF DATA. EVENT 4/4. NO INJURIES.
PART 019761R TABLET PC TREX HD MONITORING - REPAIRED. THE CUSTOMER STATES THAT THE 'TABLET IS HEATING UP TO A DANGEROUS LEVEL'. CUSTOMER MENTIONED MULTIPLE TABLETS 4 IN TOTAL. THE TABLETS THEY HAVE ARE HEATING UP AND KEEP ON RESTARTING, CAUSING THE LOST OF DATA. EVENT 4/4. NO INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2072695 | TABLET PC TREX HD MONITORING - REPAIRED | TABLET PC TREX HD MONITORING - REPAIRED | GWQ | NATUS MEDICAL INCORPORATED | 019761R | 025269471553 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |