FDA Adverse Event Malfunction Summary report: N

EXACTAMED ORAL SYRINGES

MDR report key: 1755833 · Received June 18, 2010

Report

Report Number
MW5016668
Event Type
Malfunction
Date Received
June 18, 2010
Date of Event
June 4, 2010
Report Date
June 18, 2010
Manufacturer
EXACTAMED
Product Code
KYX
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

AS PER OUR HOSPITAL PROCEDURE, PATIENT RECEIVED CLEAR SYRINGES -EXACTAMED FROM BAXA- WITH THE BLACK MARKS ON THE LINE OF THE SYRINGE WHERE THE CAREGIVERS WERE TO DRAW UP THE ORAL MEDICATION. THE LINE WAS APPROPRIATELY DELINEATED ON THE "TIC" MARK FOR 3.75 ML; HOWEVER, THE LINE CROSSED THE NUMBER "4" WHICH WAS DELINEATED TO THE SIDE OF THE TIC MARKS. IN FACT, THE NUMBERS DENOTING MILLILITERS 1, 2, 3, 4, AND 5 DID NOT CORRESPOND TO THE APPROPRIATE "TIC" MARKS ON THE SYRINGE. THE FATHER BROUGHT THIS TO OUR ATTENTION AND NEW SYRINGES FROM ANOTHER COMPANY WERE GIVEN TO HIM. THERE WAS NO ADVERSE OUTCOME TO THIS PATIENT. DIAGNOSIS OR REASON FOR USE: MEDICATION ADMINISTRATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXACTAMED ORAL SYRINGES 5 ML KYX EXACTAMED

Patients

Seq Age Sex Outcome Treatment
1 3 YR