FDA Adverse Event Malfunction Summary report: N

HAMILTON MEDICAL AG

MDR report key: 17555880 · Received August 16, 2023

Report

Report Number
3001421318-2023-02893
Event Type
Malfunction
Date Received
August 16, 2023
Date of Event
July 31, 2023
Report Date
August 16, 2023
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
07630002808590
PMA / PMN Number
K201658
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG COMES TO THE CONCLUSION: FSN WAS INITIATED: Z-0267-2023 (B)(6) 2022: IN VERY RARE CASES, THE BACKLIGHT OF THE HAMILTON-C6 DISPLAY MAY FAIL. THIS IS ATTRIBUTABLE TO A RESET OF THE HMI CONTROLLER (HMI RESET) WHICH CAUSES THE DISPLAY TO FAIL FOR 2-3 SECONDS. THE DEVICE'S VENTILATION FUNCTION IS NOT AFFECTED BY THIS.

Description of Event or Problem · 0

THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: SUMMARY: INTERACTION PANEL BLACKOUT FOR 2-3 SEC (HMI RESET).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
410240 HAMILTON MEDICAL AG HAMILTON-C6 VENTILATOR CBK HAMILTON MEDICAL AG 07630002808590

Patients

Seq Age Sex Outcome Treatment
1 Unknown