COOLSCULPTING SYSTEM
Report
- Report Number
- 3007215625-2023-12756
- Event Type
- Injury
- Date Received
- August 15, 2023
- Report Date
- November 21, 2023
- Manufacturer
- ALLERGAN (PLEASANTON)
- Product Code
- OOK
- PMA / PMN Number
- K160259
- Removal / Correction Number
- Z-1347-2022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
CORRECTED DATA D1 - BRAND NAME.
THIS IS A HISTORICAL RECORD AND REFLECTS REPORTS THAT WERE RECEIVED BY THE COMPANY PRIOR TO APRIL 2021. THIS MDR IS ASSOCIATED WITH A PARALLEL PLATE APPLICATOR, THESE APPLICATORS WERE DISCONTINUED AND RECALLED IN 2022. ACCORDING TO THE COOLSCULPTING USER MANUAL, UNDER RARE ADVERSE EVENTS, PARADOXICAL ADIPOSE HYPERPLASIA (PH/PAH) IS CHARACTERIZED BY A VISIBLY ENLARGED TISSUE VOLUME WITHIN THE TREATMENT AREA, WHICH MAY DEVELOP TWO TO FIVE MONTHS AFTER TREATMENT. SURGICAL INTERVENTION MAY BE REQUIRED. PAH IS NOT RELATED TO ANY COOLSCULPTING DEVICE FAILURE MODE BUT IT IS OBSERVED MORE FREQUENTLY WITH THE PARALLEL PLATE APPLICATORS. PAH IS INCLUDED IN THE RISK MANAGEMENT FILES OF THE DEVICE AS AN INHERENT RISK TO THE USE OF CRYOLIPOLYSIS FOR LOCALIZED FAT REDUCTION.
ALLERGAN AESTHETICS RECEIVED A REPORT OF A PATIENT TREATED WITH COOLSCULPTING ON (B)(6) 2016 TO THE LOWER ABDOMEN USING A COOLCORE APPLICATOR WHO DEVELOPED PARADOXICAL HYPERPLASIA (PAH/PH) OF THE TREATED AREA ON AN UNKNOWN DATE AFTER TREATMENT, DIAGNOSED ON (B)(6) 2017. TISSUE DESCRIBED AS FIR M, SOLID AND HARDER COMPARED TO THE SURROUNDING TISSUE WHICH WAS SQUISHY. THE HARDENED AREA ALSO HAD SOME DEMARCATION OF WHERE THE APPLICATOR WAS PLACE.
ALLERGAN AESTHETICS RECEIVED A REPORT OF A PATIENT TREATED WITH COOLSCULPTING ON (B)(6) 2016 TO THE LOWER ABDOMEN USING A COOLCORE APPLICATOR WHO DEVELOPED PARADOXICAL HYPERPLASIA (PAH/PH) OF THE TREATED AREA ON AN UNKNOWN DATE AFTER TREATMENT, DIAGNOSED ON (B)(6) 2017. TISSUE DESCRIBED AS FIR M, SOLID AND HARDER COMPARED TO THE SURROUNDING TISSUE WHICH WAS SQUISHY. THE HARDENED AREA ALSO HAD SOME DEMARCATION OF WHERE THE APPLICATOR WAS PLACE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 962901 | COOLSCULPTING SYSTEM | DERMAL COOLING PACK/VACUUM/MASSAGER | OOK | ALLERGAN (PLEASANTON) | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Female | Required Intervention |