VACUETTE® TUBE 6 ML LH LITHIUM HEPARIN 13X100 GREEN CAP-BLACK RING, NON-RIDGED
Report
- Report Number
- 1125230-2023-00039
- Event Type
- Malfunction
- Date Received
- August 15, 2023
- Report Date
- August 15, 2023
- Manufacturer
- GREINER BIO-ONE NA INC.
- Product Code
- JKA
- PMA / PMN Number
- K960857
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
COMPLAINT STATEMENT (B)(4): RECEIVED 9PCS 456088/B220734D FOR EVALUATION. RECEIVED CUSTOMER PICTURE. WE HAVE NO FURTHER INVENTORY OF THE MATERIAL/BATCH. WE HAVE NO FURTHER COMPLAINTS ON THE MATERIAL/BATCH. A CHECK OF QUALITY, PRODUCTION, AND MAINTENANCE RECORDS REVEALED NO DEVIATIONS IN RELATION TO THE REPORTED ERROR. SAMPLES WERE TESTED, ACCORDING TO GBO STANDARD TESTING PROCEDURES, WITH REGARDS TO CORRECT ASSEMBLY AND DRAW VOLUME ACCORDING TO ISO 6710 'SINGLE USE CONTAINERS FOR VENOUS BLOOD SPECIMEN COLLECTION' AND CLSI GP39-A6 REGULATIONS FOR 'EVACUATED TUBES AND ADDITIVES FOR BLOOD SPECIMEN COLLECTION'. TUBES WERE VERIFIED TO BE CORRECTLY ASSEMBLED WITH NO DEVIATIONS OBSERVED. BOTH STANDARDS SPECIFY THE DRAW VOLUME TO BE WITHIN +/- 10% RANGE OF THE NOMINAL FILL VOLUME. NO VISUAL DEVIATION IN FILL VOLUME, SPORADIC LOW OR HIGH FILL, COULD BE OBSERVED IN THE TESTED SAMPLES. ALL TUBES TESTED FILLED WITHIN THE +/-10% TOLERANCE RANGE. THE ALLEGED MALFUNCTION COULD NOT BE DUPLICATED.
CUSTOMER REPORTED THE TUBES ARE NOT FILLING COMPLETELY. THE CUSTOMER PROVIDED A PHOTO WHICH THEY ADVISED IS AN UNDERFILLED TUBE; APPROXIMATELY 1.3ML BELOW THE FILL MARK. THE CUSTOMER ADVISED TO FILL THE TUBE COMPLETELY THEY MUST COLLECT THE BLOOD WITH A SYRINGE AND FORCE FILL THE LAST 1ML OF BLOOD. THE CUSTOMER REPORTED THE INCIDENT OCCURRED SEVERAL TIMES WHILE THEY FOLLOW ALL PHLEBOTOMY BEST PRACTICES AND ARE PERFORMING VENIPUNCTURES ON HEALTHY PATIENT POPULATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1246865 | VACUETTE® TUBE 6 ML LH LITHIUM HEPARIN 13X100 GREEN CAP-BLACK RING, NON-RIDGED | EVACUATED BLOOD COLLECTION TUBE | JKA | GREINER BIO-ONE NA INC. | 456088 | B220734D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |