FDA Adverse Event Malfunction Summary report: N

VACUETTE® TUBE 6 ML LH LITHIUM HEPARIN 13X100 GREEN CAP-BLACK RING, NON-RIDGED

MDR report key: 17551020 · Received August 15, 2023

Report

Report Number
1125230-2023-00039
Event Type
Malfunction
Date Received
August 15, 2023
Report Date
August 15, 2023
Manufacturer
GREINER BIO-ONE NA INC.
Product Code
JKA
PMA / PMN Number
K960857
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

COMPLAINT STATEMENT (B)(4): RECEIVED 9PCS 456088/B220734D FOR EVALUATION. RECEIVED CUSTOMER PICTURE. WE HAVE NO FURTHER INVENTORY OF THE MATERIAL/BATCH. WE HAVE NO FURTHER COMPLAINTS ON THE MATERIAL/BATCH. A CHECK OF QUALITY, PRODUCTION, AND MAINTENANCE RECORDS REVEALED NO DEVIATIONS IN RELATION TO THE REPORTED ERROR. SAMPLES WERE TESTED, ACCORDING TO GBO STANDARD TESTING PROCEDURES, WITH REGARDS TO CORRECT ASSEMBLY AND DRAW VOLUME ACCORDING TO ISO 6710 'SINGLE USE CONTAINERS FOR VENOUS BLOOD SPECIMEN COLLECTION' AND CLSI GP39-A6 REGULATIONS FOR 'EVACUATED TUBES AND ADDITIVES FOR BLOOD SPECIMEN COLLECTION'. TUBES WERE VERIFIED TO BE CORRECTLY ASSEMBLED WITH NO DEVIATIONS OBSERVED. BOTH STANDARDS SPECIFY THE DRAW VOLUME TO BE WITHIN +/- 10% RANGE OF THE NOMINAL FILL VOLUME. NO VISUAL DEVIATION IN FILL VOLUME, SPORADIC LOW OR HIGH FILL, COULD BE OBSERVED IN THE TESTED SAMPLES. ALL TUBES TESTED FILLED WITHIN THE +/-10% TOLERANCE RANGE. THE ALLEGED MALFUNCTION COULD NOT BE DUPLICATED.

Description of Event or Problem · 0

CUSTOMER REPORTED THE TUBES ARE NOT FILLING COMPLETELY. THE CUSTOMER PROVIDED A PHOTO WHICH THEY ADVISED IS AN UNDERFILLED TUBE; APPROXIMATELY 1.3ML BELOW THE FILL MARK. THE CUSTOMER ADVISED TO FILL THE TUBE COMPLETELY THEY MUST COLLECT THE BLOOD WITH A SYRINGE AND FORCE FILL THE LAST 1ML OF BLOOD. THE CUSTOMER REPORTED THE INCIDENT OCCURRED SEVERAL TIMES WHILE THEY FOLLOW ALL PHLEBOTOMY BEST PRACTICES AND ARE PERFORMING VENIPUNCTURES ON HEALTHY PATIENT POPULATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1246865 VACUETTE® TUBE 6 ML LH LITHIUM HEPARIN 13X100 GREEN CAP-BLACK RING, NON-RIDGED EVACUATED BLOOD COLLECTION TUBE JKA GREINER BIO-ONE NA INC. 456088 B220734D

Patients

Seq Age Sex Outcome Treatment
1 Unknown