FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,13

MDR report key: 17549437 · Received August 15, 2023

Report

Report Number
0002023141-2023-02252
Event Type
Injury
Date Received
August 15, 2023
Date of Event
June 11, 2023
Report Date
January 22, 2024
Manufacturer
L
Product Code
DZE
PMA / PMN Number
K072589
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. D9: DEVICE AVAILABILITY AND RETURN DATE WERE UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODE WAS ADDED: 3331 AND 4110. H10: NARRATIVE/DATA WAS UPDATED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 63794506. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. STERILIZATION RECORD ST# (B)(4) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 63794506 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATIONS, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMVIE CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATIONS, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENTS REMAIN NON-VERIFIABLE AS THEY ARE A MEDICAL CONDITION. MONTHLY TRENDING TO DATE HAS NOT IDENTIFIED ANY STATISTICAL TRIGGERS SUGGESTING POTENTIAL DESIGN OR MANUFACTURING RELATED ISSUES. PREVIOUSLY COMPLETED INVESTIGATIONS FOR INFECTION, BONE LOSS, AND/OR PERI-IMPLANTITIS PROBLEMS REPORTED IN ASSOCIATION WITH IMPLANT FAILURE HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES IMPACTING THE MANUFACTURING AND STERILIZATION PROCESSES. THEREFORE, ESCALATION TO CAPA IS NOT REQUIRED.

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). A1: PATIENT IDENTIFIER - (B)(6). A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NON-EXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. H3 OTHER TEXT: SUMMARY INVESTIGATION.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT LOOSENING WAS FOUND AT THE IMPLANT SITES AT TOOTH LOCATIONS #19 AND #20. PERI-IMPLANTITIS AND BONE RESORPTION WERE FOUND AND THE IMPLANTS WERE REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1793815 IMP,TSV,4.1MM,SBM,13 DENTAL IMPLANT DZE L 63794506

Patients

Seq Age Sex Outcome Treatment
1 34 YR Female Required Intervention