FDA Adverse Event Death Summary report: N

ALLEGIANCE HEALTH CARE AIRLIFE INTUBATION ADAPTER

MDR report key: 175456 · Received June 30, 1998

Report

Report Number
175456
Event Type
Death
Date Received
June 30, 1998
Date of Event
June 23, 1998
Report Date
June 25, 1998
Manufacturer
ALLEGIANCE HEALTH CARE
Product Code
BZA
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
PA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

UNABLE TO VENTILATE PT. THE AIRLIFE INTUBATION ADAPTER USED AS THE CONNECTOR FOR THE CIRCUIT EXTENSION FROM THE VENTILATOR WAS DEFECTIVE. THE ADAPTER WAS FOUND TO HAVE A PLASTIC MEMBRANE TOTALLY OCCLUDING THE AIRFLOW TO THE INLET SIDE OF THE VENTILATOR. PT SUFFERED CEREBRAL ANOXIA WHICH MAY HAVE RESULTED IN PT'S DEMISE. THIS REPORT IS A FOLLOW-UP OF NO. 3902560000-1998-0015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALLEGIANCE HEALTH CARE AIRLIFE INTUBATION ADAPTER AIRLIFE INTUBATION ADAPTER BZA ALLEGIANCE HEALTH CARE * Y8C1917

Patients

Seq Age Sex Outcome Treatment
1 20 MO Death| H| L| O| R| S NARKOMED 2B ANESTHESIA MACHINE.