FDA Adverse Event Malfunction Summary report: N

BN PROSPEC SYSTEM

MDR report key: 17539626 · Received August 14, 2023

Report

Report Number
9610806-2023-00011
Event Type
Malfunction
Date Received
August 14, 2023
Date of Event
July 20, 2023
Report Date
August 14, 2023
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Product Code
JZW
UDI-DI
00630414002736
PMA / PMN Number
K001647
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN OUTSIDE OF THE UNITED STATES (OUS) CUSTOMER CONTACTED A SIEMENS CUSTOMER CARE CENTER (CCC) AND REPORTED THAT A FALSELY DEPRESSED N LATEX CYSTATIN C (N CYSC) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A BN PROSPEC SYSTEM. QUALITY CONTROL AND CALIBRATION WERE ACCEPTABLE ON THE DAY OF EVENT. THE CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER'S SITE. DURING THE VISIT, THE CSE OBSERVED THAT THE PROBE WAS MISPOSITIONED. THE CSE CORRECTED THE PROBE ALIGNMENT. THEN, THE CSE RAN QUALITY CONTROLS, WHICH RECOVERED WITHIN ACCEPTABLE RANGES. SIEMENS EVALUATED THE INSTRUMENT DATA AND DETERMINED THAT THERE WERE NO HARDWARE ERRORS GENERATED BY THE INSTRUMENT. THE CAUSE OF THE FALSELY DEPRESSED N CYSC RESULT IS UNKNOWN. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.

Description of Event or Problem · 0

A FALSELY DEPRESSED N LATEX CYSTATIN C (N CYSC) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A BN PROSPEC SYSTEM. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULT. ON THE FOLLOWING DAY, THE SAMPLE WAS REPROCESSED ON THE ORIGINAL INSTRUMENT TWICE. THE REPEAT RESULTS WERE REPORTED AS CORRECT RESULTS TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO A FALSELY DEPRESSED N CYSC RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1084104 BN PROSPEC SYSTEM BN PROSPEC SYSTEM JZW SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH BN PROSPEC SYSTEM 00630414002736

Patients

Seq Age Sex Outcome Treatment
1 Unknown