FDA Adverse Event Malfunction Summary report: N

110" 15 DROP PRIMARY SET W/2 CLAVE¿ REMV 2 GANG 4-WAY STOPCOCKS,2 CHECK VALVES,

MDR report key: 17511229 · Received August 11, 2023

Report

Report Number
9617594-2023-00535
Event Type
Malfunction
Date Received
August 11, 2023
Date of Event
July 1, 2023
Report Date
July 18, 2023
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
FMG
UDI-DI
00887709012523
PMA / PMN Number
K964435
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT OF DEFECTIVE/MALFUNCTIONING COMPONENTS ON ITEM B90111 COULD NOT BE CONFIRMED BY INVESTIGATION. SINCE NO PRODUCT SAMPLES, PICTURES, OR VIDEOS WERE RECEIVED FOR INVESTIGATION. WITHOUT THE RETURN OF THE USED SAMPLE A COMPREHENSIVE FAILURE INVESTIGATION CANNOT BE PERFORMED AND A CAUSE CANNOT BE DETERMINED. THE DEVICE HISTORY REVIEW (DHR) FOR LOT 6050629 WAS REVIEWED AND NO NON CONFORMITIES WERE FOUND THAT WOULD HAVE LED TO THE REPORTED COMPLAINT.

Description of Event or Problem · 0

THE EVENT OCCURRED ON AN UNSPECIFIED DATE AND INVOLVED A 110" 15 DROP PRIMARY SET W/2 CLAVE¿ REMV 2 GANG 4-WAY STOPCOCKS,2 CHECK VALVES, ROTATING LUER, 1 EXT, WHERE IT WAS REPORTED THE INTRAVENOUS (IV) TUBE IS NOT WORKING PROPERLY. THE CUSTOMER STATED THAT THEY CAN PRIME THE TUBING, BUT ONCE HOOKED UP TO A PATIENT IT DOES NOT RUN CORRECTLY. FURTHERMORE, IF THE CUSTOMER TAKES THE FIRST PART OF THE TUBING OFF AT THE STOPCOCK, IT SEEMS TO WORK, BUT THEY NEED THIS PIECE ATTACHED. THE CUSTOMER REPORTED THAT THE SEPARATION/ DISCONNECTION OCCURS IF LATTER PART OF STOPCOCK IS DISCONNECTED. IT WAS REPORTED THAT TUBING HAS BEEN HOOKED UP TO PATIENTS AND THEN HAD TO BE CHANGED. THERE WAS PATIENT INVOLVEMENT, BUT NO REPORT OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
763278 110" 15 DROP PRIMARY SET W/2 CLAVE¿ REMV 2 GANG 4-WAY STOPCOCKS,2 CHECK VALVES, STOPCOCK, I.V. SET FMG ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 6050629 00887709012523

Patients

Seq Age Sex Outcome Treatment
1 Unknown LACTATED RINGER'S, UNK MFR