PUMP MMT-1781K 670G V4.11 MG
Report
- Report Number
- 2032227-2023-253683
- Event Type
- Malfunction
- Date Received
- August 11, 2023
- Date of Event
- March 3, 2023
- Report Date
- August 11, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000187415
- PMA / PMN Number
- P160017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- 003
Narratives
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. PUMP RECEIVED WITH A CRACKED RETAINER RING. PUMP PASSED THE DISPLACEMENT TEST AND SELF-TEST. UNIT WAS SUCCESSFULLY DOWNLOADED USING THUS FOR HISTORY FILES AND TRACES. NO ALERTS/ALARMS/ANOMALIES NOTED DURING TESTING. PUMP ERROR 3 WAS FOUND IN THE HISTORY FILES ON 03/03/2023 16:40:13.000. PUMP ERROR 3 ALARMED DUE TO A PUMP ERROR 54. PUMP ERROR 54 (ESF#3010978; FILE NUMBER: 32122; LINE NUMBER:1421) OCCURRED ON 03/03/2023 16:40:11.000 IN THE HISTORY FILES DUE TO A SOFTWARE ERROR. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE ON THE PCBA 1, PCBA 2, FORCE SENSOR AND MOTOR HOME SWITCH. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING, HOWEVER, A CRACKED RETAINER RING WAS NOTED DURING INSPECTION. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CORRODED BATTERY TUBE, BATTERY TUBE THREADS - CRACKED, SCRATCHED CASE, STAINED KEYPAD OVERLAY, CRACKED CASE (BATTERY TUBE), CRACKED CASE-CORNER OF BELT CLIP RAILS, PILLOWING KEYPAD OVERLAY, LABEL DAMAGE (SERIAL NUMBER LABEL STAINED AND FADED; END CAP ADDRESS LABEL FADED) AND CRACKED RETAINER. PUMP ERROR 3 WAS CONFIRMED DUE TO A PUMP ERROR 54. PUMP ERROR 54 WAS CONFIRMED DUE TO A SOFTWARE ERROR. CRACKED RETAINER RING DAMAGE WAS CONFIRMED. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE PUMP ALARMED WITH ERROR 3 AND ERROR 54. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED; HOWEVER, THE CUSTOMER WILL DISCONTINUE USE OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1355246 | PUMP MMT-1781K 670G V4.11 MG | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1781K | HG3EPDZ | 000000763000187415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR | Unknown |