PUMP MMT-1714K 630G BLACK MMOL CANADA
Report
- Report Number
- 2032227-2023-253303
- Event Type
- Malfunction
- Date Received
- August 11, 2023
- Date of Event
- July 23, 2023
- Report Date
- November 2, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- UDI-DI
- 000000763000315702
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- 003
Narratives
NO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM NOTED DURING BOOT UP. THE PUMP PASSED THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DAT AT 0.08700 INCHES. NO UNEXPECTED CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM NOTED DURING THE TESTING. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. SUCCESSFULLY DOWNLOADED COMLINK3 FILE. THERE WERE NO FATAL CRITICAL ALARMS, OR THREE CONSECUTIVE CRITICAL HANDLING ALARMS FOUND IN THE FORMATTED HISTORY FILE. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM WAS GENERATED DUE TO USART TEST FAILURE IN THE INTERN BOOTLOADER (CODE 0X00000046), SUSPECTING THE HW. PLEASE SEE BELOW FOR PUMP ERRORS/ALARMS NOTED 2 DAYS PRIOR TO THE EVENT DATE 23-JUL-2023 IN THE FORMATTED HISTORY FILE. PUMP ERROR 3 ALARM (FILE NUMBER: 2002 LINE NUMBER: 2803) WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/22/2023 09:26:34.000. PUMP ERROR 3 ALARM (FILE NUMBER: 2002 LINE NUMBER: 2803) WAS CONFIRMED, SUSPECTED ON HW. PUMP ERROR 4 ALARM (FILE NUMBER: 32122 LINE NUMBER: 2727) WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/22/2023 09:26:34.000. PUMP ERROR 4 ALARM (FILE NUMBER: 32122 LINE NUMBER: 2727) WAS CONFIRMED, SUSPECTED ON HW. PUMP ERROR 68 ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/22/2023 09:26:36.000, 07/22/2023 09:27:24.000. PUMP ERROR 49 ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/22/2023 09:26:36.000, 07/22/2023 09:26:57.000, 07/22/2023 09:27:24.000, 07/22/2023 09:27:37.000. PUMP ERROR 23 ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/22/2023 09:27:10.000, 07/22/2023 09:27:54.000. PUMP ERROR 68 ALARM, PUMP ERROR 49 ALARM AND PUMP ERROR 23 ALARM WERE EXPECTED SINCE THE PUMP RESTART DUE TO PUMP ERROR 3 ALARM/PUMP ERROR 4 ALARM. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND CORROSION ON THE PCBA 1 AND PCBA 2. NO CORROSION OR MOISTURE DAMAGE FOUND ON THE FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A MISSING DISPLAY WINDOW/COVER, A CRACKED KEYPAD OVERLAY, A PILLOWING KEYPAD OVERLAY, A SCRATCHED CASE, A SERIAL NUMBER LABEL MISSING AND A BROKEN BELT CLIP RAILS. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM WAS CONFIRMED DUE TO CORROSION ON THE PCBA 1 AND PCBA 2. ALSO, PUMP ERROR 3 ALARM (FILE NUMBER: 2002 LINE NUMBER: 2803) AND PUMP ERROR 4 ALARM (FILE NUMBER: 32122 LINE NUMBER: 2727) WERE CONFIRMED ACCORDING IN THE FORMATTED HISTORY FILE DUE TO CORROSION ON THE PCBA 1 AND PCBA 2. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED OUTSIDE THE UNITED STATES. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER RECEIVED A OPEN BOOK IMAGE ALARM ON THE INSULIN PUMP SCREEN. TROUBLESHOOTING WAS PERFORMED AND THE ISSUE WAS NOT RESOLVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE INSULIN PUMP AND THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1072773 | PUMP MMT-1714K 630G BLACK MMOL CANADA | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1714K | HG46JRT | 000000763000315702 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR | Unknown |