FDA Adverse Event Malfunction Summary report: N

BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II)

MDR report key: 17507397 · Received August 11, 2023

Report

Report Number
1119779-2023-00871
Event Type
Malfunction
Date Received
August 11, 2023
Date of Event
July 26, 2023
Report Date
October 17, 2023
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JSG
PMA / PMN Number
PREAMENDMENT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: 251239 #3143197 . WE COULD CONFIRM THIS ISSUE AS A REPORT FROM RETURNED SAMPLE AND PHOTO. THIS ISSUE WAS CONTAMINATION. NO ISSUE IN RETENTION SAMPLE. NO ISSUE IN DHR. NO TREND. ROOT CAUSE WAS UNDETERMINED. WE WILL CONTINUE TO MONITOR THIS LOT. THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTED AND/OR ADDITIONAL INFORMATION. D9: DEVICE AVAILABLE FOR EVALUATION: YES. D9: RETURNED TO MANUFACTURER ON: 12-OCT-2023 H3 OTHER TEXT : SEE H.10.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE WITH BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) "BLACK" FOREIGN MATTER WAS DISCOVERED IN THE MEDIA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: ACCORDING TO THE CUSTOMER'S REPORT, BLACK SPOTS WERE FOUND IN THE MEDIA AND THE AFFECTED MEDIA WAS THUS NOT USED.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE WITH BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) "BLACK" FOREIGN MATTER WAS DISCOVERED IN THE MEDIA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: ACCORDING TO THE CUSTOMER'S REPORT, BLACK SPOTS WERE FOUND IN THE MEDIA AND THE AFFECTED MEDIA WAS THUS NOT USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1072749 BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL JSG BECTON DICKINSON & CO. (SPARKS) 3143197

Patients

Seq Age Sex Outcome Treatment
1 Unknown