BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II)
Report
- Report Number
- 1119779-2023-00871
- Event Type
- Malfunction
- Date Received
- August 11, 2023
- Date of Event
- July 26, 2023
- Report Date
- October 17, 2023
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- JSG
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION SUMMARY: 251239 #3143197 . WE COULD CONFIRM THIS ISSUE AS A REPORT FROM RETURNED SAMPLE AND PHOTO. THIS ISSUE WAS CONTAMINATION. NO ISSUE IN RETENTION SAMPLE. NO ISSUE IN DHR. NO TREND. ROOT CAUSE WAS UNDETERMINED. WE WILL CONTINUE TO MONITOR THIS LOT. THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTED AND/OR ADDITIONAL INFORMATION. D9: DEVICE AVAILABLE FOR EVALUATION: YES. D9: RETURNED TO MANUFACTURER ON: 12-OCT-2023 H3 OTHER TEXT : SEE H.10.
H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT PRIOR TO USE WITH BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) "BLACK" FOREIGN MATTER WAS DISCOVERED IN THE MEDIA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: ACCORDING TO THE CUSTOMER'S REPORT, BLACK SPOTS WERE FOUND IN THE MEDIA AND THE AFFECTED MEDIA WAS THUS NOT USED.
IT WAS REPORTED THAT PRIOR TO USE WITH BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) "BLACK" FOREIGN MATTER WAS DISCOVERED IN THE MEDIA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: ACCORDING TO THE CUSTOMER'S REPORT, BLACK SPOTS WERE FOUND IN THE MEDIA AND THE AFFECTED MEDIA WAS THUS NOT USED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1072749 | BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) | CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL | JSG | BECTON DICKINSON & CO. (SPARKS) | 3143197 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |