ALLURA XPER FD
Report
- Report Number
- 3003768277-2023-04361
- Event Type
- Malfunction
- Date Received
- August 10, 2023
- Date of Event
- July 27, 2023
- Report Date
- February 11, 2025
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product Code
- OWB
- UDI-DI
- 00884838059085
- PMA / PMN Number
- K130638
- Removal / Correction Number
- 3003768277-06/05/2023-00
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
PHILIPS HAS INVESTIGATED THIS COMPLAINT. A PHILIPS SERVICE ENGINEER INSPECTED THE SYSTEM ONSITE AND CONFIRMED THAT 4 LOOSE BOLTS WERE FOUND, AND A METAL STRIP IS MISSING. THE METAL STRIP PREVENTS PLAY BETWEEN THE METAL BLOCK AND THE LINEAR BEARING, AND ADDITIONAL SHEAR STRESS TO THE BOLTS. THE C-ARC IS CONNECTED TO THE METAL BLOCK. UPON SOME FUNCTIONAL TEST FSE ARRANGED A NEW METAL STRIP WAS. THE FSE INSTALLED THE METAL STRIP AND MOUNTED THE BOLTS WITH LOCTITE, AFTER INSTALLING THE FLEXARM METAL STRIP, THE SYSTEM WAS RETURNED TO USE IN GOOD WORKING ORDER. THE CODES WERE UPDATED BASED ON THE INVESTIGATION OUTCOME.
THE REPORTED EVENT OCCURRED PRIOR TO COMPLETION OF THE FIELD CORRECTION 3003768277-06/05/2023-003-C, WHICH ADDRESSES THE CAUSES ASSOCIATED WITH THE REPORTED MALFUNCTION.
IT HAS BEEN REPORTED TO PHILIPS THAT THE ALLURA XPER FD20 OR TABLE SYSTEM FLEXMOVE TRANSVERSAL BEARING WAS LOOSE. THE SYSTEM WAS NOT IN CLINICAL USE AND NO HARM HAS BEEN REPORTED. UPON EVALUATION IT WAS DETERMINED THAT THE WRONG STRIP WAS ON THE FLEXMOVE COMPONENT. THE STRIP WAS REPLACED, AND THE SYSTEM WAS RETURNED TO FUNCTIONALITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 797635 | ALLURA XPER FD | INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM | OWB | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. | ALLURA XPER FD20 OR TABLE | 00884838059085 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |