630G INSULIN PUMP MMT-1715KL 630G
Report
- Report Number
- 2032227-2023-252749
- Event Type
- Malfunction
- Date Received
- August 10, 2023
- Date of Event
- July 22, 2023
- Report Date
- September 27, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- UDI-DI
- 000000763000316655
- PMA / PMN Number
- P150001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- 003
Narratives
PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. SUCCESSFULLY UTILIZED CREST AND THUS TO DOWNLOAD HISTORY FILES, TRACES AND COMLINK3 FILES. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM AND PUMP ERROR 35 ALARM CONFIRMED IN THE FORMATTED HISTORY FILE ON 07/22/2023 09:48:00.000 AND 07/22/2023 09:58:00.000. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND CORROSION ON THE PCBA 1, PCBA 2 AND FORCE SENSOR. NO CORROSION OR MOISTURE DAMAGE FOUND ON THE MOTOR AND VIBRATOR ASSEMBLY NOTED. PERFORM BRUSH CLEANING WITH THE ISOPROPYL ALCOHOL THE MOISTURE DAMAGE AREAS AND REINSTALLED THE ORIGINAL ELECTRONICS, CASE, INTERNAL BATTERY AND MOTOR. THE CRITICAL PUMP ERROR OCCURRED DURING TESTING. IN CONCLUSION, CRITICAL PUMP ERROR (OPEN BOOK) AND PUMP ERROR 35 ALARM FOUND IN THE FORMATTED HISTORY FILE DUE TO CORROSION ON THE PCBA 1, PCBA 2 AND FORCE SENSOR. PLEASE SEE BELOW FOR PUMP ERRORS/ALARMS NOTED 2 DAYS PRIOR TO THE EVENT DATE 22-JUL-2023 IN THE FORMATTED HISTORY FILE. INSERT BATTERY ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/22/2023 09:56:36.000, 07/22/2023 09:58:13.000. FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/22/2023 09:57:03.000. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. NO POWER ERROR 25, LOW BATTERY ALERT, POWER LOSS ALARM AND REPLACE BATTERY ALERT/REPLACE BATTERY NOW ALARM NOTED 2 DAYS PRIOR TO THE EVENT DATE 22-JUL-2023 IN THE FORMATTED HISTORY FILE. INSERT BATTERY ALARM WAS EXPECTED SINCE THE BATTERY WAS REMOVED FROM THE PUMP. UPON CHECKING ON THE POWER DATA/DETAIL TRACE FILE, FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS EXPECTED DUE TO PUMP BATTERY DOES NOT HAVE ENOUGH POWER. THE CUSTOMER MAY HAVE USED A NO POWER/DEPLETED BATTERY. UNABLE TO TEST FOR FAILED BATTERY ALERT/BATTERY FAILED ALARM DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS UNKNOWN. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED BATTERY TUBE THREADS, A SCRATCHED CASE, A CRACKED CASE-CORNER OF BELT CLIP RAILS NEAR THE BATTERY TUBE COMPARTMENT AND A SERIAL NUMBER LABEL FADING/PEELING. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO FATAL ALARM PUMP ERROR 35. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM AND PUMP ERROR 35 ALARM CONFIRMED DUE TO CORROSION ON THE PCBA 1, PCBA 2 AND FORCE SENSOR. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER RECEIVED A OPEN BOOK IMAGE ALARM. TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE INSULIN PUMP PERFORMED SAFETY CHECKS AND THE ERROR WAS FOUND. IT WAS FOUND THAT THE PUMP HAD AN OPEN-BOOK IMAGE ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE INSULIN PUMP AND WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1758908 | 630G INSULIN PUMP MMT-1715KL 630G | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1715KL | HG6APMW | 000000763000316655 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Female |