PUMP MMT-1880L MM770G BLE MG
Report
- Report Number
- 2032227-2023-252732
- Event Type
- Malfunction
- Date Received
- August 10, 2023
- Date of Event
- July 21, 2023
- Report Date
- October 20, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000439866
- PMA / PMN Number
- P160017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE PUMP WAS RECEIVED WITH A CRACKED CASE BEHIND THE PUMP NEAR THE BATTERY TUBE COMPARTMENT. THE PUMP PASSED THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DAT AT 0.08735 INCHES. THE PUMP WAS MONITORED AND NO UNEXPECTED ERROR MESSAGE (UNABLE TO DETECT ACTIVE INSULIN) NOTED. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUMP. SUCCESSFULLY UPLOADED PUMP TO CARELINK. IN FURTHER FULL REVIEW OF THE PUMP HISTORY/TRACES ON THE EVENT DATE OF 21-JUL-2023, THERE IS NO UNEXPECTED ALARMS/SUSPENDS. PLEASE SEE BELOW FOR THE DAILY TOTAL OF BASAL/BOLUS AND ALL INSULIN DELIVERED ON THE EVENT DATE 21-JUL-2023 LISTED ON SMARTSOLVE. DAILYTOTALOFALLINSULINDELIVERED: 446250 (44.625 U). DAILYTOTALOFBASALINSULINDELIVERED: 255250 (25.525 U). DAILYTOTALOFBOLUSINSULINDELIVERED: 191000 (19.1 U). 07/21/2023 00:21:25.000 NORMALBOLUSDELIVERED (220). BOLUSPROGRAMMINGMETHOD: CL1BGCORRECTION (6). NORMALBOLUSAMOUNTPROGRAMMED: 14000 (1.4 U). BOLUSAMOUNTDELIVERED: 14000 (1.4 U). 07/21/2023 10:07:39.000 NORMALBOLUSDELIVERED (220). BOLUSPROGRAMMINGMETHOD: CL1FOODBOLUS (7). NORMALBOLUSAMOUNTPROGRAMMED: 70000 (7 U). BOLUSAMOUNTDELIVERED: 70000 (7 U). 07/21/2023 12:33:02.000 NORMALBOLUSDELIVERED (220). BOLUSPROGRAMMINGMETHOD: BOLUSWIZARD (1). NORMALBOLUSAMOUNTPROGRAMMED: 65000 (6.5 U). BOLUSAMOUNTDELIVERED: 65000 (6.5 U). 07/21/2023 13:39:18.000 NORMALBOLUSDELIVERED (220). BOLUSPROGRAMMINGMETHOD: BOLUSWIZARD (1). NORMALBOLUSAMOUNTPROGRAMMED: 8000 (0.8 U). BOLUSAMOUNTDELIVERED: 8000 (0.8 U). 07/21/2023 14:50:31.000 NORMALBOLUSDELIVERED (220). BOLUSPROGRAMMINGMETHOD: CL1BGCORRECTION (6). NORMALBOLUSAMOUNTPROGRAMMED: 1000 (0.1 U). BOLUSAMOUNTDELIVERED: 1000 (0.1 U). 07/21/2023 19:59:45.000 NORMALBOLUSDELIVERED (220). BOLUSPROGRAMMINGMETHOD: CL1FOODBOLUS (7). NORMALBOLUSAMOUNTPROGRAMMED: 33000 (3.3 U). BOLUSAMOUNTDELIVERED: 33000 (3.3 U). THE PUMP WAS PROGRAMMED WITH MULTIPLE BOLUS DELIVERIES AND ALL BOLUS DELIVERED PROPERLY THEIR INDICATED AMOUNTS (AT QUICK BOLUS SPEED) AND WERE PROPERLY RECORDED IN THE DAILY HISTORY. NO BOLUS DELIVERY ANOMALY OR HISTORY ANOMALY NOTED. NO UNDER DELIVERY ANOMALY OR OVER DELIVERY ANOMALY NOTED DURING TESTING. PLEASE SEE BELOW FOR PUMP ERRORS/ALARMS NOTED 1 WEEK PRIOR TO THE EVENT DATE 21-JUL-2023 IN THE FORMATTED HISTORY FILE. LOSTSENSOR1ALERT (780) WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/16/2023 04:03:00.000, 07/16/2023 08:28:00.000, 07/21/2023 09:55:00.000. LOSTSENSOR2ALERT (781) WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/21/2023 10:11:00.000. THE PUMP WAS PROGRAMMED WITH A TEST GUARDIAN LINK (3) TRANSMITTER AND A 240 MG/DL GLUCOSE SENSOR SIMULATOR. THE PUMP CONNECTED SUCCESSFULLY TO THE TRANSMITTER AND DISPLAYED ¿TRANSMITTER CONNECTION SUCCESSFUL¿. THE PUMP COMMUNICATED PROPERLY WITH GLUCOSE SENSOR SIMULATOR AND DISPLAYED THE CALIBRATE YOUR SENSOR ALARM PROPERLY AFTER COMPLETION OF THE WARM UP. THE PUMP CALIBRATED AND DISPLAYED THE PROGRAMMED VALUE OF 239 MG/DL PROPERLY ON THE DISPLAY GRAPH. NO LOST SENSOR ALERT OR UNEXPECTED SENSOR ERRORS OR ANOMALIES WERE NOTED DURING TESTING. LOW BATTERY ALERT WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/21/2023 09:55:00.000. INSERT BATTERY ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 07/21/2023 13:46:00.000. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. NO ALARMS/ALERTS WERE NOTED. INSERT BATTERY ALARM WAS EXPECTED SINCE THE BATTERY WAS REMOVED FROM THE PUMP. EARLIEST POWER DATA AVAILABLE PER THE POWER MANAGEMENT TOOL/DETAIL TRACE FILE IS ON 02-AUG-2023 AT 3:30:57 AM. THERE WAS NO POWER DATA AVAILABLE FOR THE EVENT DATE OF 21-JUL-2023. UNABLE TO CHECK POWER DATA FOR LOW BATTERY ALERT. NO UNEXPECTED LOW BATTERY ALERT NOTED DURING TESTING. THE TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY AND A SCRATCHED CASE. COSMETIC DAMAGE WAS CONFIRMED AT THE BACK (TO FRONT) OF THE PUMP. THE PUMP PASSED ALL THE REQUIRED TESTING. CUSTOMER ALLEGED FOR POSSIBLE UNDER/OVER DELIVERY ANOMALY AND UNEXPECTED ERROR MESSAGE (UNABLE TO DETECT ACTIVE INSULIN) WERE NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER STATED THAT THE INSULIN PUMP ALERTED AND WAS UNABLE TO DETECT ACTIVE INSULIN, AS WELL AS A CRACK IN THE INSULIN PUMP FROM THE REAR TO THE FRONT. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL CONTINUE USING THE INSULIN PUMP AND WILL NOT BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 950436 | PUMP MMT-1880L MM770G BLE MG | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1880L | HG60AVP | 000000763000439866 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Female |