POMPE À PERFUSION PLUM 360¿, COMPATIBLE AVEC LE LOGICIEL ICU MEDICAL MEDNET¿
Report
- Report Number
- 9615050-2023-00310
- Event Type
- Malfunction
- Date Received
- August 9, 2023
- Date of Event
- August 1, 2023
- Report Date
- August 1, 2023
- Manufacturer
- ICU MEDICAL COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K161469
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE CUSTOMER'S COMPLAINT WAS INVESTIGATED AND TESTING WAS PERFORMED ON THE PUMP. THE LOGS WERE REVIEWED. TIME PERIOD: 21-JUN-2023 TO 25-JUL-2023, SEVEN ENTRIES OF N185 (PEAK PROXIMAL OCCLUSION A NON-DELIVERY) WERE IDENTIFIED ON 25-JULY-2023. THE PROBABLE CAUSE FOR THE IDENTIFIED ERROR CODES COULD NOT BE DETERMINED. ALTHOUGH THE PUMP FAILED VISUAL INSPECTION DUE TO A CRACKED REAR CASE/DOOR ASSEMBLY, THE DEVICE PASSED OCCLUSION TESTING WITHOUT ISSUE. THE PRESSURE DETECTOR WILL BE RECALIBRATED AS A PRECAUTION DURING REPAIR. D9 - DATE RETURNED TO MFG: 14AUG2023.
THE COMPLAINT INVESTIGATION WAS UPDATED ON 03OCT2023 TO INCLUDE THE FOLLOWING INFORMATION: IN ADDITION TO THE THE DEVICE PASSING THE OCCLUSION TESTING WITHOUT ISSUE DURING THE COMPLAINT INVESTIGATION, THE INFUSION PUMP DEVICE ALSO PASSED ALL TESTING, INCLUDING A FULL PERFORMANCE VERIFICATION TESTING (PVT) WITHOUT ANY ISSUE. THE DEVICE ALARMED AUDIBLY WHEN EXPECTED DURING OCCLUSION TESTS. THE COMPLAINT COULD NOT BE CONFIRMED THROUGH TESTING OR THROUGH ALARM LOGS REVIEW. NO PROBABLE CAUSE FOR THE CUSTOMER'S REPORTED ISSUE COULD BE DETERMINED.
THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION; HOWEVER, IT HAS NOT YET BEEN RECEIVED.
THE EVENT OCCURRED ON AN UNSPECIFIED DATE AND INVOLVED A POMPE À PERFUSION PLUM 360¿, COMPATIBLE AVEC LE LOGICIEL ICU MEDICAL MEDNET¿. THE FOLLOWING ISSUE WAS REPORTED BY THE CUSTOMER: ¿MISSED OCCLUSION¿. THERE WAS NO HUMAN HARM REPORTED AS A RESULT OF THIS EVENT/COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1904393 | POMPE À PERFUSION PLUM 360¿, COMPATIBLE AVEC LE LOGICIEL ICU MEDICAL MEDNET¿ | PUMP, INFUSION | FRN | ICU MEDICAL COSTA RICA LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |