FDA Adverse Event Malfunction Summary report: N

BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WITH BLOOD

MDR report key: 17485198 · Received August 8, 2023

Report

Report Number
1710034-2023-00891
Event Type
Malfunction
Date Received
August 8, 2023
Date of Event
July 25, 2023
Report Date
September 15, 2023
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825332
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: OUR QUALITY ENGINEER INSPECTED THE PHOTOGRAPH SUBMITTED FOR EVALUATION. BD RECEIVED ONE PHOTOGRAPH WHICH DISPLAYED A 20GX1.00IN INSYTE AUTOGUARD FROM LOT NUMBER 3033950. THE CATHETER WAS FULL OF BLOOD RESIDUE AND THE NEEDLE WAS PROTRUDING THROUGH THE SIDE OF THE CATHETER TUBING. DUE TO THE EVIDENCE OF USE, THE DAMAGE LIKELY OCCURRED DURING THE INSERTION PROCESS. THE IFU INDICATES TO NOT REINSERT THE NEEDLE INTO THE CATHETER DURING THE INSERTION PROCESS AS DAMAGE MAY OCCUR. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH.

Additional Manufacturer Narrative · 0

E1. ADDRESS INFORMATION WAS NOT ABLE TO BE OBTAINED, THEREFORE, NJ WAS USED AS A PLACE HOLDER. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 5 OF THE BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WITH BLOOD CONTROL EXPERIENCED NEEDLE THROUGH CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE HAS PUNCTURED THROUGH THE CATHETER, 5 OCCURRENCES AS REPORTED BY CUSTOMER.

Description of Event or Problem · 0

IT WAS REPORTED THAT 5 OF THE BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WITH BLOOD CONTROL EXPERIENCED NEEDLE THROUGH CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE HAS PUNCTURED THROUGH THE CATHETER, 5 OCCURRENCES AS REPORTED BY CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1688349 BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WITH BLOOD INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 3033950 00382903825332

Patients

Seq Age Sex Outcome Treatment
1 Unknown