ICL (IMPLANTABLE COLLAMER LENS)
Report
- Report Number
- 2023826-2010-00665
- Event Type
- Injury
- Date Received
- July 1, 2010
- Report Date
- June 4, 2010
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- MTA
- PMA / PMN Number
- P030016
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B) (4): EVALUATION CONCLUSION: AT THE CONCLUSION OF THE ORIGINAL INVESTIGATION OPENED FOR THE ISSUE OF ICL VAULTING (BOTH INADEQUATE AND EXCESSIVE), IT WAS DETERMINED THAT THE MOST LIKELY ROOT CAUSE FOR BOTH INADEQUATE AND EXCESSIVE VAULT IS DIFFICULTY IN CLINICAL SIZING. IN NO CASE HAS A RETURNED LENS EXHIBITED A LENGTH MEASUREMENT OUTSIDE THE EXPECTED RANGE ACCORDING TO THE LABELED LENGTH. PRE-OPERATIVE MEASUREMENT TECHNOLOGY AVAILABLE TO CLINICIANS IN THE PAST DID NOT PROVIDE A DIRECT MEASUREMENT OF THE SULCUS, TO WHICH THE ICL SHOULD BE SIZED. THE CURRENT PRACTICE IN SELECTING AN APPROPRIATE ICL FOR A PATIENT IS TO MEASURE WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH AND, THROUGH CORRELATIVE ALGORITHMS, PREDICT THE LENGTH OF THE ICL THAT WILL BRIDGE THE SULCUS. THIS METHOD OF SIZING BASED UPON WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH MEASUREMENTS WAS USED IN THE FDA CLINICAL TRIAL AND IS RECOMMENDED IN THE CURRENT LABELING OF THE ICL. IF THE PREDICTED LENGTH IS TOO SHORT, THE RESULT MAY BE AN INADEQUATE VAULT. IT THE PREDICTED LENGTH IS TOO LONG, THE RESULT MAY BE AN EXCESSIVE VAULT. (B) (4).
IT WAS REPORTED THAT THE SURGEON INSERTED AN ICM130V4 13.0MM IMPLANTABLE COLLAMER LENS AND EXCESSIVE VAULT WAS NOTED, WHICH IS ASSOCIATED WITH NARROWING OF THE ANGLE AND SHALLOWING OF THE ACD. PATIENT EXPERIENCED ELEVATED IOP. THE LENS REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ICL (IMPLANTABLE COLLAMER LENS) | INTRAOCULAR LENS | MTA | STAAR SURGICAL CO. | ICM130V4 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE: MODEL UNK, LOT NUMBER UNK| FOAM TIP PLUNGER: MODEL UNK, LOT NUMBER UNK| INJECTOR: MODEL UNK, LOT NUMBER UNK |