LENSX LASER SYSTEM
Report
- Report Number
- 2028159-2023-01104
- Event Type
- Malfunction
- Date Received
- August 8, 2023
- Date of Event
- July 7, 2023
- Report Date
- October 11, 2023
- Manufacturer
- ALCON LENSX, INC.
- Product Code
- OOE
- UDI-DI
- 00380659981623
- PMA / PMN Number
- K163551
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
REASSESSMENT IN THE FDA PATIENT EVENT CODES (1). BLURRED VISION WAS NOT QUALIFYING FOR REPORTABLE EVENT BECAUSE POST-OPERATIVE TREATMENT RESULT WAS UNKNOWN. HENCE, BLURRED VISION HAS BEEN REMOVED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
A MANUFACTURING DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. SERVICE HISTORY WAS REVIEWED FOR THE SYSTEM. NO SERVICE RECORD RELEVANT TO THE COMPLAINT REPORTED EVENT WAS FOUND. HOWEVER, THE SYSTEM WAS LAST SERVICED PRIOR TO THE REPORTED EVENT PER SERVICE RECORD (SR) OPENED. THE SYSTEM FOUND TO MEET ALL COSMETIC AND PERFORMANCE STANDARDS. BASED ON THE INFORMATION OBTAINED, THE ROOT CAUSE OF THE REPORTED EVENT IS INCONCLUSIVE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ALCON LENSX (SITE (B)(4)) IS NO LONGER OPERATIONAL. LENSX MANUFACTURED PRODUCTS ARE MAINTAINED AND INVESTIGATED BY THE ALCON RESEARCH, LTD. IRVINE TECHNOLOGY CENTER SITE (B)(4)). THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
A HEALTHCARE PROFESSIONAL REPORTED THAT THE BLURRED VISION ON FIRST DAY POST OPERATIVE AND FOUND STRIKES ON FLAP, SURGEON LIFTED THE FLAP AND CLEAR IN THE LEFT EYE OF A PATIENT, DURING LASIK SURGERY. THERE ARE TWO RELATED REPORTS FOR THIS PATIENT. THIS REPORT ADDRESSES THE PATIENT LEFT EYE AND ANOTHER MANUFACTURER REPORT WILL BE FILED FOR THE FELLOW EYE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1261573 | LENSX LASER SYSTEM | OPHTHALMIC FEMTOSECOND LASER | OOE | ALCON LENSX, INC. | 550 | NA | 00380659981623 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | Male |