PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2953144-2010-01232
- Event Type
- Injury
- Date Received
- July 1, 2010
- Date of Event
- June 7, 2010
- Report Date
- June 7, 2010
- Manufacturer
- ABBOTT VASCULAR - REDWOOD CITY
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
(B) (4). (B) (4). DEVICE #2: PART # 12673-05, LOT #790020-6H INDICATED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED.
DEVICE #1 ISSUE: NEEDLE-TO-CUFF MISS. TIME OF DEVICE ISSUE. DURING VESSEL CLOSURE. ADVERSE EVENT: FAILURE TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, WHEN THE NEEDLE PLUNGER WAS REMOVED, NO SUTURES WERE PRESENT. A SECOND PROGLIDE DEVICE WAS ATTEMPTED BUT ALSO RESULTED IN NO SUTURES PRESENT. A THIRD PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | MGB | ABBOTT VASCULAR - REDWOOD CITY | NA | 79020-6H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | DEVICE #2: PROGLIDE, PART #12673-05, LOT #79020-6H |