FDA Adverse Event
Death
Summary report: N
INVACARE RELIANT 450
MDR report key: 1747436
·
Received February 10, 2010
Report
- Report Number
- 1747436
- Event Type
- Death
- Date Received
- February 10, 2010
- Date of Event
- January 28, 2010
- Report Date
- February 2, 2010
- Manufacturer
- INVACARE
- Product Code
- FSA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
ON (B)(6) 2010, RESIDENT WAS BEING TRANSFERRED VIA INVACARE RELIANT LIFT. WHEN THE BOTTOM OF THE SLING BECAME SLACK AND DISLODGED FROM HANGER. RESIDENT SLID FROM SLING, FALLING TO FLOOR RESULTING IN HEMATOMA TO RIGHT SIDE OF HEAD. LIFT TAKEN OUT OF SERVICE. SLING TAKEN OUT OF SERVICE. RESIDENT SENT TO ER ON (B)(6) 2010 AND RETURNED WITH NO SERIOUS INJURY. RESIDENT CTB ON (B)(6) 2010.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INVACARE RELIANT 450 | NONE | FSA | INVACARE | RPL-450-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 93 YR | Death |