FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36

MDR report key: 17458348 · Received August 4, 2023

Report

Report Number
3005075853-2023-05487
Event Type
Malfunction
Date Received
August 4, 2023
Date of Event
June 12, 2023
Report Date
August 4, 2023
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
UDI-DI
10705036014621
PMA / PMN Number
K132612
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). DATE SENT: 8/4/2023. D4 BATCH #: X9603L. INVESTIGATION SUMMARY THE PRODUCT WAS RETURNED FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE DEVICE WAS RETURNED WITH THE TISSUE PAD DETACHED AND NOT RETURNED. IN ADDITION THE BLADE WAS DAMAGED THE DEVICE WAS CONNECTED TO A TEST HAND PIECE AND A GEN11, AND THE DEVICE DID ACTIVATE DURING FUNCTIONAL TESTING. NO UNEXPECTED READY TO PRE-RUN TEST ISSUE WAS OBSERVED AT ANY TIME AS REPORTED. THE DEVICE WAS DISASSEMBLED TO INSPECT THE INTERNAL COMPONENTS AND NO ANOMALIES WERE FOUND. BASED ON THE CONDITION OF THE TISSUE PAD, A PROBABLE CAUSE OF THIS DAMAGE IS THAT THE CLAMP OF THE DEVICE MAY HAVE BEEN CLOSED AND THE INSTRUMENT ACTIVATED WITHOUT TISSUE PRESENT. CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. KEEP THE CLAMP ARM OPEN WHEN BACK CUTTING OR WHILE THE BLADE IS ACTIVE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. THE RESULTING DAMAGE CONTRIBUTES TO THE REMOVAL OF THE PAD FROM THE CLAMP ARM. THE CLEANING OF THE PAD, NOT IN ACCORDANCE WITH THE IFU, CAN ALSO RESULT IN REMOVAL OF THE PAD DURING USE. AS PART OF THE QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT/BATCH X9603L, AND NO NON-CONFORMANCES WERE IDENTIFIED. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A GASTRO-INTESTINAL PROCEDURE THE SCALPEL PASSED FIRST PRE-RUN TEST. AFTER WORKING SEVERAL MINUTES, THE SCALPEL WAS REQUIRED TO PRE-RUN BY THE SYSTEM AGAIN. THE SCALPEL COULD NOT WORK UNINTERRUPTED. CHANGED TO ANOTHER DEVICE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
875684 HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36 INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. X9603L 10705036014621

Patients

Seq Age Sex Outcome Treatment
1 Unknown HANDPIECE AND GENERATOR