HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Report
- Report Number
- 3005075853-2023-05487
- Event Type
- Malfunction
- Date Received
- August 4, 2023
- Date of Event
- June 12, 2023
- Report Date
- August 4, 2023
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- UDI-DI
- 10705036014621
- PMA / PMN Number
- K132612
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). DATE SENT: 8/4/2023. D4 BATCH #: X9603L. INVESTIGATION SUMMARY THE PRODUCT WAS RETURNED FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE DEVICE WAS RETURNED WITH THE TISSUE PAD DETACHED AND NOT RETURNED. IN ADDITION THE BLADE WAS DAMAGED THE DEVICE WAS CONNECTED TO A TEST HAND PIECE AND A GEN11, AND THE DEVICE DID ACTIVATE DURING FUNCTIONAL TESTING. NO UNEXPECTED READY TO PRE-RUN TEST ISSUE WAS OBSERVED AT ANY TIME AS REPORTED. THE DEVICE WAS DISASSEMBLED TO INSPECT THE INTERNAL COMPONENTS AND NO ANOMALIES WERE FOUND. BASED ON THE CONDITION OF THE TISSUE PAD, A PROBABLE CAUSE OF THIS DAMAGE IS THAT THE CLAMP OF THE DEVICE MAY HAVE BEEN CLOSED AND THE INSTRUMENT ACTIVATED WITHOUT TISSUE PRESENT. CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. KEEP THE CLAMP ARM OPEN WHEN BACK CUTTING OR WHILE THE BLADE IS ACTIVE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. THE RESULTING DAMAGE CONTRIBUTES TO THE REMOVAL OF THE PAD FROM THE CLAMP ARM. THE CLEANING OF THE PAD, NOT IN ACCORDANCE WITH THE IFU, CAN ALSO RESULT IN REMOVAL OF THE PAD DURING USE. AS PART OF THE QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT/BATCH X9603L, AND NO NON-CONFORMANCES WERE IDENTIFIED. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING A GASTRO-INTESTINAL PROCEDURE THE SCALPEL PASSED FIRST PRE-RUN TEST. AFTER WORKING SEVERAL MINUTES, THE SCALPEL WAS REQUIRED TO PRE-RUN BY THE SYSTEM AGAIN. THE SCALPEL COULD NOT WORK UNINTERRUPTED. CHANGED TO ANOTHER DEVICE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 875684 | HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36 | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | X9603L | 10705036014621 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | HANDPIECE AND GENERATOR |