FDA Adverse Event
Malfunction
Summary report: N
OT PING METER
MDR report key: 1745633
·
Received July 2, 2010
Report
- Report Number
- 2939301-2010-05297
- Event Type
- Malfunction
- Date Received
- July 2, 2010
- Report Date
- June 15, 2010
- Manufacturer
- LIFESCAN, INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
Description of Event or Problem · 1
A PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "259, 555, 100, 459, 130 AND 120 MG/DL" WITH A LIFESCAN METER AND "344, 484, 254 254, 264 AND 261 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <= 30% AND/OR <= 30 MG/DL. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |