FDA Adverse Event Malfunction Summary report: N

LENSX LASER SYSTEM

MDR report key: 17452413 · Received August 3, 2023

Report

Report Number
2028159-2023-01096
Event Type
Malfunction
Date Received
August 3, 2023
Date of Event
June 23, 2023
Report Date
October 12, 2023
Manufacturer
ALCON LENSX, INC.
Product Code
OOE
UDI-DI
00380659981623
PMA / PMN Number
K163551
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3., H.6.: INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ALCON LENSX (SITE #3008772169) IS NO LONGER OPERATIONAL. LENSX MANUFACTURED PRODUCTS ARE MAINTAINED AND INVESTIGATED BY THE ALCON RESEARCH, LTD. IRVINE TECHNOLOGY CENTER SITE #2028159). THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

A MANUFACTURING DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. SERVICE HISTORY WAS REVIEWED FOR THE SYSTEM. NO SERVICE RECORD RELEVANT TO THE COMPLAINT REPORTED EVENT WAS FOUND. HOWEVER, THE SYSTEM WAS LAST SERVICED PRIOR TO THE REPORTED EVENT PER SERVICE RECORD (SR) OPENED. THE SYSTEM FOUND TO MEET ALL COSMETIC AND PERFORMANCE STANDARD. THE CUSTOMER REPORTED EVENT WAS NOT CONFIRMED. THUS, BASED ON THE INFORMATION OBTAINED, THE ROOT CAUSE OF THE REPORTED EVENT IS INCONCLUSIVE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED THAT THE FLAP SIDE CUT WAS INTERRUPTED IN THE UNKNOWN EYE OF A PATIENT, DURING SURGERY. THE NEW PATIENT INTERFACE WAS USED BUT THE FLAP CUT DIDN¿T HAPPEN PROPERLY. THE SURGEON COMPLETED THE SURGERY WITH MANUAL SEPARATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
284687 LENSX LASER SYSTEM OPHTHALMIC FEMTOSECOND LASER OOE ALCON LENSX, INC. 550 NA 00380659981623

Patients

Seq Age Sex Outcome Treatment
1 Unknown