VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE
Report
- Report Number
- 9610773-2023-02116
- Event Type
- Malfunction
- Date Received
- August 2, 2023
- Report Date
- August 5, 2024
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- HET
- UDI-DI
- 04042761074971
- PMA / PMN Number
- K111788
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE COULD NOT BE ESTABLISHED. THE SUGGESTED PHENOMENON OF "GREEN IMAGE" WAS CONFIRMED, AND PRESUMED TO HAVE BEEN DUE TO THE FOLLOWING: 1. CABLE PINS OF ODU CONNECTOR [ODU BEING A COMPANY'S NAME] ARE TOO SMALL FOR SHIELD CABLES ¿ SHIELD CABLES ARE GROUPED OF UP TO FOUR SINGLE CABLES AND THIS SHIELD GROUP IS TOO BIG IN DIAMETER FOR THE PINS OF ODU CONNECTOR. SOLDERING JOINTS OF SHIELD GROUP MIGHT BE TOO WEAK TO WITHSTAND THE FORCES WITHIN CABLE. 2. SEALING COMPOUND WITHIN ODU CONNECTOR DOES NOT SEAL THE SOLDERING JOINTS AND BARE CABLE CONTACTS COMPLETELY AGAINST STEAM ¿ CORROSION ON THE SOLDERING JOINTS AND BARE CABLES CAN BE SEEN AFTER SEVERAL AUTOCLAVE CYCLES. 3. MANUFACTURING JIG (B)(4) NOT FULLY SUITABLE FOR SOLDERING PROCESS ¿ CABLES AND SOLDERING JOINTS ARE UNDER STRESS DURING MANUFACTURING PROCESS AND THIS CAN LEAD TO PREMATURE DAMAGES OF THE SOLDERING JOINTS. 4. SOLDERING PROCESS AT THE MANUFACTURER IS NOT UP TO DATE. NEW GUIDELINES FOR SOLDERING PROCESS ARE AVAILABLE ¿ THE NEW GUIDELINES IMPROVE SOLDERING STABILITY AND MANUFACTURABILITY OF ¿GOOD¿ SOLDERING JOINTS. THE REPORTED EVENT GREEN IMAGE (CAUSED BY CABLE UNIT): ACTIONS WERE INITIATED FOR THIS ERROR PATTERN. AS PART OF THIS CHANGE, THE MANUFACTURING PROCESS REGARDING THE SOLDERED JOINTS BETWEEN CABLE AND ODU PLUG WAS IMPROVED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
THE DEVICE WAS RETURNED TO AN OLYMPUS REPAIR FACILITY, AND AN EVALUATION OF THE DEVICE WAS PERFORMED. DURING EVALUATION, THE CUSTOMERS¿ INITIAL REPORT OF GREEN IMAGE WAS CONFIRMED. THEY ALSO NOTED MULTIPLE DENTS ON THE OUTER TUBE AND A LOOSE LIGHT GUIDE ADAPTER. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS PROVIDED BY THE USER FACILITY.
A USER FACILITY REPORTED TO OLYMPUS THE IMAGE IS GREEN WHILE USING VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE. THE MALFUNCTION WAS FOUND AT PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1337522 | VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE | ENDOEYE HD II | HET | OLYMPUS WINTER & IBE GMBH | WA50042A | 04042761074971 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |