RESTORE ADVNACED RECHARGEABLE
Report
- Report Number
- 3004209178-2010-05143
- Event Type
- Malfunction
- Date Received
- June 30, 2010
- Date of Event
- March 1, 2010
- Report Date
- June 2, 2010
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO., JUNCOS
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4)
PATIENT STATES STIMULATION IS INTERMITTENT. ACCORDING TO THE PATIENT PROGRAMMER, STIMULATION IS ALWAYS ON. HE SAID THERE ARE PERIODS OF TIME WHERE HE DOESN'T FEEL STIMULATION UNTIL HE ADJUSTS THE PARAMETERS. DEVICE IS NORMALLY AT 3.20V. THE RECHARGE INTERVAL HAS INCREASED WITH THIS NEW SYSTEM. OLD SYSTEM AVERAGED FROM 20-25 AND NEW SYSTEM IS 30-40 DAYS. REPORTS COUPLING IS GOOD AND RECHARGES IN AN AVERAGE OF 3 HOURS WHEN THE BATTERY IS 1/4 FULL. HISTORY OF PAST RECHARGE SESSIONS: (B)(6)2009 -- 39 DAYS, (B)(6)2010 -- 32 DAYS, (B)(6)2010 -- 22 DAYS, (B)(6)2010 -- 33 DAYS AND (B)(6)2010 -- 55 DAYS. PATIENT STATES HE USED TO BE ABLE TO WALK NINE HOLES OF GOLF BUT NOW IS NOT ABLE TO DUE TO THE PAIN. PATIENT WILL KEEP A HISTORY OF THE STIMULATION ON/OFF AND ACTIVITY AT THE TIME TO DETERMINE IF RECHARGING INTERVALS ARE APPROPRIATE. MEDTRONIC REPRESENTATIVE REPORTS AT THE LAST INS INTERROGATION, TWO CONTACTS HAD >10,000 OHMS AND SHE WAS ABLE TO CHANGE THE PROGRAM TO REMOVE THOSE. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF RECEIVED, A FOLLOW UP REPORT WILL BE SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESTORE ADVNACED RECHARGEABLE | LGW | MDT PUERTO RICO OPERATIONS CO., JUNCOS | 37713 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | IMPLANTED:| ACCESSORY: MODEL 37752, LOT# NKA123723N| PROGRAMMER: MODEL 37743, LOT# NKE119993N| IMPLANTED:| EXPLANTED:| LEAD: MODEL 3776, LOT# V006142| EXPLANTED:| LEAD: MODEL 3778, LOT# V177008013| IMPLANTABLE:| LEAD: MODEL 3776, LOT# V006142| EXPLANTED: |