FDA Adverse Event Malfunction Summary report: N

BD POSIFLUSH¿ SF SALINE SYRINGE

MDR report key: 17442493 · Received August 2, 2023

Report

Report Number
9616657-2023-00029
Event Type
Malfunction
Date Received
August 2, 2023
Date of Event
June 30, 2023
Report Date
August 23, 2023
Manufacturer
BECTON DICKINSON DE MEXICO
Product Code
NGT
UDI-DI
00382903065530
PMA / PMN Number
K153481
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 306553 AND LOT NUMBER 2243740. THE REVIEW DID NOT REVEAL ANY DETECTED NON-CONFORMANCES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS INCIDENT. TO AID IN THE INVESTIGATION OF THIS ISSUE, FOUR (4) SAMPLES WERE RETURNED FOR EVALUATION BY OUR QUALITY TEAM. TWO (2) OF THE BLISTER PACKAGES WERE SEPARATED AND OPENED WITH THE SYRINGES STILL INSIDE. THE REMAINING TWO (2) SYRINGES WERE SEPARATED, AND THE BLISTER PACKAGES WERE OPENED. THROUGH THE INSPECTION, WE CAN CONFIRM THAT THE MOST PROBABLE CAUSE FOR THIS ISSUE WAS FIBER TEARING. A SMALL AMOUNT OF FIBER TEARING NATURALLY OCCURS WHEN SEPARATING INDIVIDUAL BLISTERS AND WHEN SEPARATING THE PAPER TOP WEB FROM THE PLASTIC BOTTOM WEB.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD POSIFLUSH¿ SF SALINE SYRINGE PACKAGING TORE WHILE OPENING AND LEFT PAPER SHARDS IN THE STERILE FIELD. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTED THE CHADS FROM THE PERFORATION ARE FLAKING ONTO THE STERILE FIELD. WHEN THEY'RE OPENED FROM PAPER PACK, SOME OF THE PAPER SEEMS TO BE FLAKING OFF ONTO THE STERILE FIELD. THEY ARE NOT SURE THE EXACT AMOUNT SO FAR THAT HAVE HAD THE ISSUE. IS THERE ANY ADVERSE EVENT OCCURRED? NO IS THERE ANY PATIENT HARM, INJURY, OR NEGATIVE OUTCOME THAT HAS NOT ALREADY BEEN DISCUSSED? NO".

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD POSIFLUSH¿ SF SALINE SYRINGE PACKAGING TORE WHILE OPENING AND LEFT PAPER SHARDS IN THE STERILE FIELD. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTED THE CHADS FROM THE PERFORATION ARE FLAKING ONTO THE STERILE FIELD. WHEN THEY'RE OPENED FROM PAPER PACK, SOME OF THE PAPER SEEMS TO BE FLAKING OFF ONTO THE STERILE FIELD. THEY ARE NOT SURE THE EXACT AMOUNT SO FAR THAT HAVE HAD THE ISSUE... -IS THERE ANY ADVERSE EVENT OCCURRED? NO. -IS THERE ANY PATIENT HARM, INJURY, OR NEGATIVE OUTCOME THAT HAS NOT ALREADY BEEN DISCUSSED? NO".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1385058 BD POSIFLUSH¿ SF SALINE SYRINGE SALINE, VASCULAR ACCESS FLUSH NGT BECTON DICKINSON DE MEXICO 2243740 00382903065530

Patients

Seq Age Sex Outcome Treatment
1 Unknown