PRIDE MOBILITY PRODUCTS

MDR report key: 17436024 · Received August 1, 2023

Report

Report Number
2530130-2023-00118
Event Type
Injury
Date Received
August 1, 2023
Date of Event
June 26, 2023
Report Date
June 26, 2024
Manufacturer
PRIDE MOBILITY PRODUCTS
Product Code
ITI
UDI-DI
00606509200158
PMA / PMN Number
K143383
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PER CURTIS: CTLDC1540 THE CONTROLLER WAS LOW POWER BENCH TESTED WITH THE 1741 HANDCONTROL THAT CAME WITH IT. THE CONTROLLER PASSED LOW POWER BENCH TESTING. NO ERRACTIC BEHAVIOR AS REPORTED. FAULT HISTORY: JOYSTICK OOC, CURRENT SENSOR FAULT, JOYSTICK OUT OF RANGE, BRAKE ON FAULT, JOYSTICK STUCK OOC, HANDCONTROL EEROM FAULT. AWAITING TO BE TESTED ON THE ATS... UPDATE 11/08/2023: THE CONTROLLER WAS TESTED AND PASSED ATS. THE CONTROLLER WAS NFF. CTLDC1554 THE JOYSITCK KNOB (CPN (B)(4)) WAS MISSING FROM THE HANDCONTROL. THE HANDCONTROL WAS LOW POWER BENCH TESTED WITH THE 1740 CONTROLLER THAT CAME WITH IT. THE HANDCONTROL PASSED LOW POWER BENCH TESTING. NO ERRACTIC BEHAVIOR AS REPORTED. AWAITING TO BE TESTED ON THE ATS...UPDATE 11/08/2023: THE HANDCONTROL WAS TESTED AND PASSED ATS. THE HANDCONTROL WAS NFF.

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT YET BEEN MADE AVAILABLE FOR EVALUATION. SHOULD FURTHER INFORMATION OR THE DEVICE BECOME AVAILABLE, A FOLLOW-UP REPORT WILL THEN BE ISSUED.

Description of Event or Problem · 0

ALLEGES POWER CHAIR WAS SPINNING FAST AT HIGH VELOCITY CLOCKWISE BEFORE THE CHAIR AND RESIDENT ALLEGEDLY FELL FORWARD.

Description of Event or Problem · 0

ALLEGES POWER CHAIR WAS SPINNING FAST AT HIGH VELOCITY CLOCKWISE BEFORE THE CHAIR AND RESIDENT ALLEGEDLY FELL FORWARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2336091 PRIDE MOBILITY PRODUCTS WHEELCHAIR, POWERED ITI PRIDE MOBILITY PRODUCTS Q6 EDGE 3 N/A 00606509200158

Patients

Seq Age Sex Outcome Treatment
1 NA Male Other