FDA Adverse Event Malfunction Summary report: N

UNIVERSAL SLINGB 450 QRH

MDR report key: 17433851 · Received August 1, 2023

Report

Report Number
8030916-2023-00055
Event Type
Malfunction
Date Received
August 1, 2023
Date of Event
July 21, 2023
Report Date
August 1, 2023
Manufacturer
LIKO AB
Product Code
FSA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER PROVIDED A PICTURE OF THE BROKEN COMPOSITE SLING BAR HOOK. THE PICTURE CONFIRMS A FRACTURE IN THE COMPOSITE HOOK. THE UNIVERSAL SLINGBAR CAN BE USED IN COMBINATION WITH LIKO SLINGS THAT ATTACHES TO THE SLING BAR USING SLING LOOPS. LIKO´S UNIVERSAL SLING BARS ARE WITH TWO ASSEMBLY OPTIONS, FIXED ASSEMBLY OR WITH QUICK-RELEASE HOOK. HILLROM INVESTIGATION ON PREVIOUS SIMILAR CASES CONCLUDES THAT THE UNIVERSAL SLINGBAR WILL NOT BREAK IF IT IS USED AS INTENDED. THE UNIVERSAL SLINGBAR CAN HOLD THE LOAD IT IS DESIGNED FOR IF THE LOAD IS APPLIED ACCORDING TO INSTRUCTION FOR USE. TESTS DEMONSTRATE THAT THE COMPOSITE HOOKS ON THE UNIVERSAL SLINGBAR CAN WITHSTAND MORE THAN 780 KG LOAD UNTIL BREAKAGE. OUR INVESTIGATION FURTHER SHOWS THAT THE BREAKAGE OF THE COMPOSITE HOOK CAN OCCUR AT LOWER WEIGHT IF THE LOAD IS NOT CENTERED, THE LOAD IS APPLIED IN ONLY ONE HOOK OR THE LOAD IS NOT APPLIED INSIDE THE HOOK AS INTENDED. IN THE INSTRUCTION FOR USE FOR UNIVERSAL SLING BARS (7EN160185 REV 5), THE SAFETY INSTRUCTION SECTION STATES: - BEFORE LIFTING, ALWAYS MAKE SURE THAT THE SLING¿S STRAP LOOPS ARE CORRECTLY FASTENED TO THE SLING BAR HOOKS WHEN THE SLING STRAP IS EXTENDED, BUT BEFORE THE PATIENT IS LIFTED FROM THE UNDERLYING SURFACE. FOR SAFETY, LOAD MUST BE APPLIED TO ALL SLING BAR HOOKS ON THE SLING BAR DURING THE LIFT! ALTHOUGH THERE WAS NO PATIENT INJURY ASSOCIATED WITH THE REPORTED EVENT, THE REPORT OF THE SLING BAR HOOK BREAKING DURING PATIENT LIFT COULD CONTRIBUTE TO A SERIOUS INJURY OR DEATH, IF THE MALFUNCTION WERE TO RECUR. THEREFORE, HILLROM CONSIDERS THIS COMPLAINT A REPORTABLE MALFUNCTION.

Description of Event or Problem · 0

IT WAS REPORTED THAT ONE OF THE SLINGBAR'S HOOK BROKE. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1634702 UNIVERSAL SLINGB 450 QRH OVERHEAD TRACK PATIENT LIFTING/TRANSFER SYSTEM FSA LIKO AB 3156085

Patients

Seq Age Sex Outcome Treatment
1 Unknown