BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2023-00976
- Event Type
- Malfunction
- Date Received
- August 1, 2023
- Date of Event
- May 30, 2023
- Report Date
- July 24, 2024
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. ADDITIONALLY, BD WAS UNABLE TO DETERMINE THE SPECIFIC LOT NUMBER ASSOCIATED WITH THIS COMPLAINT; THEREFORE, A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE CONDUCTED. THIS COMPLAINT IS UNABLE TO BE CONFIRMED. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THIS COMPLAINT WILL BE REOPENED TO ASSESS THE LEVEL OF INVESTIGATION NEEDED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. D.4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H.4. DEVICE MANUFACTURE DATE: UNKNOWN. E.1. INITIAL REPORTER ADDR 1 : (B)(6).
IT WAS REPORTED WHEN USING THE BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES CUSTOMER STATES THAT THEY SEEN A LARGE INCREASE IN THE INCIDENCE OF SODIUM FLYERS ON THE PST TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: WE HAVE SEEN A LARGE INCREASE IN THE INCIDENCE OF SODIUM FLYERS ON THE VITROS IN OUR REGIONAL LABS USING PST TUBES. QUIDEL-ORTHO HAVE INVESTIGATED AND SUGGEST THE ISSUE COULD BE RELATED TO THE VACUTAINERS ¿ EITHER THE CENTRIFUGATION SPEEDS OR THE GEL INTEGRITY.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES CUSTOMER STATES THAT THEY SEEN A LARGE INCREASE IN THE INCIDENCE OF SODIUM FLYERS ON THE PST TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: WE HAVE SEEN A LARGE INCREASE IN THE INCIDENCE OF SODIUM FLYERS ON THE VITROS IN OUR REGIONAL LABS USING PST TUBES. QUIDEL-ORTHO HAVE INVESTIGATED AND SUGGEST THE ISSUE COULD BE RELATED TO THE VACUTAINERS ¿ EITHER THE CENTRIFUGATION SPEEDS OR THE GEL INTEGRITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1554255 | BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |