FDA Adverse Event Injury Summary report: N

FREESTYLE LIBRELINK

MDR report key: 17431180 · Received July 31, 2023

Report

Report Number
2954323-2023-32902
Event Type
Injury
Date Received
July 31, 2023
Date of Event
July 14, 2023
Report Date
July 31, 2024
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
PZE
PMA / PMN Number
P160030
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS COMPLAINT WAS INVESTIGATED. ADC HAS IDENTIFIED A SOFTWARE DEFECT FOR THE FREESTYLE LIBRELINK APPLICATION FOR IOS, VERSION 2.10.0, IN WHICH THE APPLICATION UPGRADE WAS UNSUCCESSFUL. THIS RESULTED IN A PERIOD WHERE USERS MAY NOT HAVE RECEIVED GLUCOSE RESULTS, OR MAY NOT HAVE BEEN ALERTED TO LOW OR HIGH GLUCOSE ALARMS. BASED ON THE INVESTIGATION, THE COMPLAINT IS CONFIRMED AND NO FURTHER INVESTIGATION ACTIVITIES FOR THIS INDIVIDUAL COMPLAINT ARE REQUIRED. THIS ISSUE WAS ADDRESSED IN THE FIELD BY ADC (B)(6). THE DEVICE MODEL NUMBER POPULATED IN SECTION D4 IS FOR THE FREESTYLE LIBRELINK IOS APPLICATION, ON MARKET IN THE UK. THIS IS SAME/SIMILAR TO US FREESTYLE LIBRELINK/FREESTYLE LIBRE 2 IOS APPLICATION PART NUMBER 71733-01/71926-01. THE DEVICE MANUFACTURER DATE FOR THE REPORTED SERIAL NUMBER IS UNKNOWN. THE DATE ENTERED IN SECTION H4 IS THE DATE ABBOTT DIABETES CARE BECAME AWARE OF THE EVENT. THIS SERVES AS A CORRECTION REPORT. SECTION H6 (ADVERSE EVENT PROBLEM) WAS INCORRECTLY DOCUMENTED IN THE PREVIOUS REPORT. CORRECTION HAS BEEN MADE. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

THIS COMPLAINT WAS INVESTIGATED. ADC HAS IDENTIFIED A SOFTWARE DEFECT FOR THE FREESTYLE LIBRELINK APPLICATION FOR IOS, VERSION 2.10.0, IN WHICH THE APPLICATION UPGRADE WAS UNSUCCESSFUL. THIS RESULTED IN A PERIOD WHERE USERS MAY NOT HAVE RECEIVED GLUCOSE RESULTS, OR MAY NOT HAVE BEEN ALERTED TO LOW OR HIGH GLUCOSE ALARMS. BASED ON THE INVESTIGATION, THE COMPLAINT IS CONFIRMED AND NO FURTHER INVESTIGATION ACTIVITIES FOR THIS INDIVIDUAL COMPLAINT ARE REQUIRED. THIS ISSUE WAS ADDRESSED IN THE FIELD BY ADC FA1029-2023. THE DEVICE MODEL NUMBER POPULATED IN SECTION D4 IS FOR THE FREESTYLE LIBRELINK IOS APPLICATION, ON MARKET IN THE UK. THIS IS SAME/SIMILAR TO US FREESTYLE LIBRELINK/FREESTYLE LIBRE 2 IOS APPLICATION PART NUMBER 71733-01/71926-01. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 0

AN ISSUE WAS REPORTED WITH THE ADC APPLICATION VERSION 2.10.0, IN USE WITH IPHONE 13 WITH IOS OPERATING SYSTEM VERSION 16.5.1. A CUSTOMER REPORTED THAT THEIR APPLICATION WAS DOWN AND IT WAS INDICATED THAT THE CUSTOMER RECEIVED TREATMENT. ADC HAS IDENTIFIED THAT FOLLOWING THE RELEASE OF THE FREESTYLE LIBRELINK (FSLL) APPLICATION FOR IOS, V2.10.0, ON (B)(6)2023 (11AM) IN THE UNITED KINGDOM (UK), SOME USERS HAVE EXPERIENCED A SITUATION WHERE THE APPLICATION UPGRADE IS UNSUCCESSFUL, AND AS A RESULT, THEY RECEIVE A WHITE SCREEN IN THE APPLICATION USER INTERFACE (UI). THIS ISSUE DOES NOT AFFECT USERS IN THE UNITED STATES. THIS WILL RESULT IN A PERIOD WHERE GLUCOSE READINGS AND ALARMS WILL NOT BE RECEIVED, AND HISTORICAL GLUCOSE READINGS WILL NOT BE ACCESSIBLE, AS THE PREVIOUS FSLL APPLICATION, V2.8.1, WAS NO LONGER AVAILABLE TO USERS ON THE APPLE APP STORE. THIS ISSUE WAS ADDRESSED IN THE FIELD BY ADC (B)(6) AND WAS RESOLVED IN THE FIELD BY THE RELEASE OF UPDATED FSLL IOS APPLICATION, MADE AVAILABLE IN THE APPLE APP STORE IN THE UK ON (B)(6)2023. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 0

AN ISSUE WAS REPORTED WITH THE ADC APPLICATION VERSION 2.10.0, IN USE WITH IPHONE 13 WITH IOS OPERATING SYSTEM VERSION 16.5.1. A CUSTOMER REPORTED THAT THEIR APPLICATION WAS DOWN AND IT WAS INDICATED THAT THE CUSTOMER RECEIVED TREATMENT. ADC HAS IDENTIFIED THAT FOLLOWING THE RELEASE OF THE FREESTYLE LIBRELINK (FSLL) APPLICATION FOR IOS, V2.10.0, ON 12-JUL-2023 (11AM) IN THE UNITED KINGDOM (UK), SOME USERS HAVE EXPERIENCED A SITUATION WHERE THE APPLICATION UPGRADE IS UNSUCCESSFUL, AND AS A RESULT, THEY RECEIVE A WHITE SCREEN IN THE APPLICATION USER INTERFACE (UI). THIS ISSUE DOES NOT AFFECT USERS IN THE UNITED STATES. THIS WILL RESULT IN A PERIOD WHERE GLUCOSE READINGS AND ALARMS WILL NOT BE RECEIVED, AND HISTORICAL GLUCOSE READINGS WILL NOT BE ACCESSIBLE, AS THE PREVIOUS FSLL APPLICATION, V2.8.1, WAS NO LONGER AVAILABLE TO USERS ON THE APPLE APP STORE. THIS ISSUE WAS ADDRESSED IN THE FIELD BY ADC FA1029-2023 AND WAS RESOLVED IN THE FIELD BY THE RELEASE OF UPDATED FSLL IOS APPLICATION, MADE AVAILABLE IN THE APPLE APP STORE IN THE UK ON 17-JUL-2023. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.