FDA Adverse Event Injury Summary report: N

PITT EASY ABUTMENT

MDR report key: 1742694 · Received June 30, 2010

Report

Report Number
9611820-2010-00009
Event Type
Injury
Date Received
June 30, 2010
Report Date
June 1, 2010
Manufacturer
SYBRON IMPLANT SOLUTIONS GMBH
Product Code
NHA
PMA / PMN Number
K053242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

AN ERROR WAS MADE ON THE INITIAL REPORT REGARDING THE DESCRIPTION OF THIS INCIDENT. IT WAS REPORTED THAT THE CONNECTION BETWEEN THE CROWN AND THE ABUTMENT BECAME LOOSE AND SUBSEQUENTLY FRACTURED. THIS STATEMENT IS INCORRECT. FURTHER INVESTIGATION INDICATED THAT THE FRACTURE ACTUALLY OCCURRED BETWEEN THE METAL AND CERAMIC PORTIONS OF THE ABUTMENT. AS STATED IN THE INITIAL REPORT, THE PRODUCT MET SPECIFICATIONS AND THE FRACTURE WAS LIKELY DUE TO THE TECHNIQUE THE CLINICIAN USED TO DRILL THROUGH THE CERAMIC CROWN IN ORDER TO REACH THE SCREW CHANNEL.

Additional Manufacturer Narrative · 1

IT APPEARS THAT THIS INCIDENT IS THE RESULT OF THE TECHNIQUE THE DOCTOR USED TO DRILL THROUGH THE CERAMIC CROWN IN ORDER TO REACH THE SCREW CHANNEL. THE DRILLING CAUSED THE LOOSENING AND SUBSEQUENT BREAKING OF THE CONNECTION BETWEEN THE CROWN AND THE ABUTMENT. THE DOCTOR DID NOT PROVIDE ANY ADDITIONAL INFORMATION REGARDING THE CEMENT OR CROWN MATERIAL THAT WAS USED. AN INCOMING GOODS INSPECTION WAS CONDUCTED ON THE RETURNED ABUTMENT AND INDICATED THAT THE PRODUCT MET SPECIFICATIONS.

Description of Event or Problem · 1

ON (B)(6), 2010 A DOCTOR REPORTED TO SYBRON IMPLANT SOLUTIONS (B)(4) THAT A CROWN BECAME DETACHED FROM A PITT EASY ABUTMENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S GLOBAL TECHNICAL SERVICE CENTER TO REPORT AN ALARM ON THE HOMECHOICE MACHINE DURING PRIME. THE TECHNICAL SERVICE REPRESENTATIVE HAD THE HOME PATIENT (HP) PUSH THE SPIKE INTO THE SUPPLY BAG AND TURN. PER THE HP, THE SPIKE HAD NOT PENETRATED THE SEAL OF THE SUPPLY BAG. THE HP RESUMED PRIME. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PITT EASY ABUTMENT ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS NHA SYBRON IMPLANT SOLUTIONS GMBH

Patients

Seq Age Sex Outcome Treatment
1 Other| R