OMNIPOD 5 POD
Report
- Report Number
- 3004464228-2023-19271
- Event Type
- Injury
- Date Received
- July 31, 2023
- Date of Event
- June 24, 2023
- Report Date
- July 11, 2023
- Manufacturer
- INSULET CORPORATION
- Product Code
- QFG
- UDI-DI
- 10385083000114
- PMA / PMN Number
- K203768
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
ACCORDING TO THE COMPLAINANT THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION. WE ARE UNABLE TO DETERMINE IF ANY PRODUCT CONDITION COULD HAVE CONTRIBUTED TO THE REPORTED HOSPITALIZATION, DISLODGED CANNULA AND DIABETIC KETOACIDOSIS. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS. HIGH BLOOD GLUCOSE IS A COMMON SYMPTOM FOR PEOPLE WITH DIABETES (GLUCOSE MONITORING DATA FROM PEOPLE WITH DIABETES INDICATE THAT ON AVERAGE, THEY CAN EXPERIENCE BLOOD GLUCOSE LEVELS ABOVE 250 MG/DL FOR 14-25% OF THE TIME[1][2][3]..), AND IT WOULD BE CHALLENGING TO SPECULATE ON A CAUSE FOR THE COMPLAINTS WITHOUT RECEIVING THE DEVICES BACK FOR AN ENGINEERING INVESTIGATION. [1] BECK RW, BERGENSTAL RM, CHENG P, KOLLMAN C, CARLSON AL, JOHNSON ML, RODBARD D. THE RELATIONSHIPS BETWEEN TIME IN RANGE, HYPERGLYCEMIA METRICS, AND HBA1C. J DIABETES SCI TECHNOL 2019;13:614-626. [1] WELSH JB, DERDZINSKI M, PARKER AS, PUHR S, JIMENEZ A, WALKER T. REAL-TIME SHARING AND FOLLOWING OF CONTINUOUS GLUCOSE MONITORING DATA IN YOUTH. DIABETES THER 2019;10:751-755. [1] PUHR S, DERDZINSKI M, WELSH JB, PARKER AS, WALKER T, PRICE DA. REAL-WORLD HYPOGLYCEMIA AVOIDANCE WITH A CONTINUOUS GLUCOSE MONITORING SYSTEM'S PREDICTIVE LOW GLUCOSE ALERT. DIABETES TECHNOL THER 2019;21:155-158.
IT WAS REPORTED THAT THE PATIENT WENT TO THE HOSPITAL AND WAS DIAGNOSED WITH DIABETIC KETOACIDOSIS (DKA). THE PATIENT'S BLOOD GLUCOSE (BG) VALUES REACHED 600 MG/DL. THE PATIENT HAD A HEADACHE, SLURRING SPEECH, WAS NAUSEOUS AND VOMITING. FOR TREATMENT, THE PATIENT WAS GIVEN INSULIN AND DIAGNOSTIC TESTS. THE PATIENT WAS DISCHARGED AFTER 12 HOURS. THE CANNULA WAS DISLODGED, WHILE WEARING THE POD BETWEEN 4 AND 24 HOURS ON THE ABDOMEN. THE POD WAS DISCARDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 604467 | OMNIPOD 5 POD | ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP | QFG | INSULET CORPORATION | PT-000435 | PH1U11072211 | 10385083000114 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Female | Hospitalization| R |