PUMP MMT-1780KPK 670G PATHWAY BLACK MG
Report
- Report Number
- 2032227-2023-245676
- Event Type
- Malfunction
- Date Received
- July 31, 2023
- Date of Event
- July 11, 2023
- Report Date
- October 5, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000190460
- PMA / PMN Number
- P150001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- 003
Narratives
THE PUMP PASSED THE SELF TEST, DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, ACTIVE CURRENT TEST, AND SLEEP CURRENT TEST. NO UNEXPECTED POWER LOSS ALARMS/ALERTS/ANOMALIES NOTED DURING TESTING. NO UNEXPECTED INSULIN FLOW BLOCKED ALARMS NOTED DURING TESTING. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. NO POWER ALARMS/ALERTS NOTED IN DOWNLOADED HISTORY ON EVENT DATE. THE POWER MANAGEMENT GRAPH CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. VERIFIED THE PUMP ALARMED NO DELIVERY DURING BASAL/BOLUS/PRIMING IN PUMP DOWNLOADED HISTORY NEAR EVENT DATE 21 TIMES. LISTED BELOW ARE THE DATES/TIMES OF NO DELIVERY ALARMS LISTED ON OR NEAR EVENT DATE ALONG WITH DURING BASAL/BOLUS/PRIMING: 2 NO DELIVERY ALARM DURING BOLUS: START DATE (B)(6) 2023 19:03:00.000 END DATE (B)(6) 2023 19:03:45.000 2 NO DELIVERY ALARM DURING PRIMING: START DATE (B)(6) 2023 19:06:40.000 END DATE (B)(6) 2023 19:22:17.000 1 NO DELIVERY ALARM DURING BOLUS: START DATE (B)(6) 2023 20:36:27.000 1 NO DELIVERY ALARM DURING BASAL: START DATE (B)(6) 2023 20:45:00.000 2 NO DELIVERY ALARM DURING BOLUS: START DATE (B)(6) 2023 20:03:52.000 END DATE (B)(6) 2023 22:04:28.000 1 NO DELIVERY ALARM DURING PRIMING: START DATE (B)(6) 2023 22:05:33.000 3 NO DELIVERY ALARM DURING BASAL: START DATE (B)(6) 2023 03:53:00.000 END DATE (B)(6) 2023 04:02:00.000 1 NO DELIVERY ALARM DURING BOLUS: START DATE (B)(6) 2023 06:57:00.000 2 NO DELIVERY ALARM DURING BASAL: START DATE (B)(6) 2023 09:56:00.000 END DATE (B)(6) 2023 09:57:00.000 1 NO DELIVERY ALARM DURING PRIMING: START DATE (B)(6) 2023 10:07:58.000 2 NO DELIVERY ALARM DURING BOLUS: START DATE (B)(6) 2023 12:03:40.000 END DATE (B)(6) 2023 18:03:12.000 2 NO DELIVERY ALARM DURING PRIMING: START DATE (B)(6) 2023 18:21:39.000 END DATE (B)(6) 2023 18:25:28.000 1 NO DELIVERY ALARM DURING BASAL: START DATE (B)(6) 2023 19:06:00.000 TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY, MOTOR OR FORCE SENSOR. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: PILLOWING KEYPAD OVERLAY, PEELING SERIAL NUMBER LABEL, AND CRACKED RETAINER. PUMP PASSED REQUIRED TESTING. NO FAILED BATTERY ALARMS NOTED DURING TESTING, FAILED BATT TEST NOT CONFIRMED. NO UNEXPECTED INSULIN FLOW BLOCKED ALARMS NOTED DURING TESTING. NO DELIVERY ALARM NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THE INSULIN PUMP HAD NO DELIVERY ALERTS, AND REJECTED THE BATTERIES ON OCCASION. NO ADDITIONAL DETAILS WERE GIVEN. THE CUSTOMER DECLINED THE TROUBLESHOOTING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. IT WAS UNKNOWN WHETHER THE CUSTOMER CONTINUED USING THE PUMP OR NOT AND WILL NOT BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 645360 | PUMP MMT-1780KPK 670G PATHWAY BLACK MG | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KPK | HG34DN4 | 000000763000190460 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Male |