ELEVATE¿ SPINAL SYSTEM
Report
- Report Number
- 1030489-2023-00549
- Event Type
- Malfunction
- Date Received
- July 31, 2023
- Date of Event
- July 6, 2023
- Report Date
- July 31, 2023
- Manufacturer
- WARSAW ORTHOPEDICS
- Product Code
- MAX
- UDI-DI
- 00643169430433
- PMA / PMN Number
- K172199
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
H3: PRODUCT ANALYSIS #7770723 : LOT # 0940225W VISUAL AND OPTICAL INSPECTION REVEALED THE IMPLANT WAS RETURNED DAMAGED. THE PEEK PORTION OF THE SPACER HAS BROKEN AWAY. BOTH TABS/EARS OF THE IMPLANT HAVE BEEN BROKEN AS WELL. THE RETURNED STATE OF THE IMPLANT SUGGEST THE SPACER WAS IN MORE OF COLLAPSED POSITION. THE PEEK PORTION MAY HAVE CAUGHT ON SOMETHING IN THE VERTEBRAL SPACE DURING THE IMPLANTATION PROCEDURE. THIS TYPE OF DAMAGE IS CONSISTENT WITH MATERIAL OVERLOAD. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR UNKNOWN INDICATION. IT WAS REPORTED THAT THE ELEVATE CAGE WAS CLOSED BEFORE IMPLANTATION. WHEN THE SURGEON WENT TO INSERT IT EXPANDED AND BROKE. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1965924 | ELEVATE¿ SPINAL SYSTEM | INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, | MAX | WARSAW ORTHOPEDICS | 7770723 | 0940225W | 00643169430433 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female |