FDA Adverse Event Malfunction Summary report: N

ELEVATE¿ SPINAL SYSTEM

MDR report key: 17422814 · Received July 31, 2023

Report

Report Number
1030489-2023-00549
Event Type
Malfunction
Date Received
July 31, 2023
Date of Event
July 6, 2023
Report Date
July 31, 2023
Manufacturer
WARSAW ORTHOPEDICS
Product Code
MAX
UDI-DI
00643169430433
PMA / PMN Number
K172199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS #7770723 : LOT # 0940225W VISUAL AND OPTICAL INSPECTION REVEALED THE IMPLANT WAS RETURNED DAMAGED. THE PEEK PORTION OF THE SPACER HAS BROKEN AWAY. BOTH TABS/EARS OF THE IMPLANT HAVE BEEN BROKEN AS WELL. THE RETURNED STATE OF THE IMPLANT SUGGEST THE SPACER WAS IN MORE OF COLLAPSED POSITION. THE PEEK PORTION MAY HAVE CAUGHT ON SOMETHING IN THE VERTEBRAL SPACE DURING THE IMPLANTATION PROCEDURE. THIS TYPE OF DAMAGE IS CONSISTENT WITH MATERIAL OVERLOAD. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR UNKNOWN INDICATION. IT WAS REPORTED THAT THE ELEVATE CAGE WAS CLOSED BEFORE IMPLANTATION. WHEN THE SURGEON WENT TO INSERT IT EXPANDED AND BROKE.  THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1965924 ELEVATE¿ SPINAL SYSTEM INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, MAX WARSAW ORTHOPEDICS 7770723 0940225W 00643169430433

Patients

Seq Age Sex Outcome Treatment
1 Female