HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2023-14490
- Event Type
- Malfunction
- Date Received
- July 31, 2023
- Date of Event
- April 29, 2022
- Report Date
- July 31, 2023
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 07630002800372
- PMA / PMN Number
- K201658
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT HAS BEEN REOPENED AND REVIEWED ACCORDING TO FDA FORM 483 INSPECTIONAL OBSERVATION EMS #2, EOBS2 FROM THE FDA INSPECTION CONDUCTED BETWEEN JULY 17 TO JULY 21, 2022 AT THE EMS AND BONADUZ SITES. A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS PRESUMABLY PREPARED FOR TREATMENT. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECT INFELLICUFF WHICH WAS REPLACED. THERE WAS NO PATIENT OR USER HARM.
SELF TEST FAILED. 244020. UNIT ALARMING CONSTANTLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1840831 | HAMILTON MEDICAL AG | HAMILTON-C6 | CBK | HAMILTON MEDICAL AG | HAMILTON-C6 | 07630002800372 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |