FDA Adverse Event Malfunction Summary report: N

BD FACSLYRIC¿

MDR report key: 17421343 · Received July 28, 2023

Report

Report Number
2916837-2023-00153
Event Type
Malfunction
Date Received
July 28, 2023
Date of Event
June 29, 2023
Report Date
November 29, 2023
Manufacturer
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Product Code
OYE
UDI-DI
00382906628756
PMA / PMN Number
H.10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE PMA / 510(K)#S REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH 510(K) NUMBER IS AS FOLLOWS: K170974, K201814 E.6. INITIAL REPORTER ADDRESS 1: (B)(6) H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION OF THE COMPLAINT, IT HAS BEEN DETERMINED THAT THE PREVIOUSLY SUBMITTED REPORT WAS SENT IN ERROR. CARRYOVER DID NOT OCCUR AND THERE WAS NO PATIENT SAMPLES AFFECTED, THEREFORE, THIS IS NOT CONSIDERED TO BE A REPORTABLE MALFUNCTION.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD FACS¿ SAMPLE PREP ASSISTANT III CARRYOVER BETWEEN PATIENT SAMPLES HAD OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER THAT SPILLOVER SEEN IN SAMPLES. 1. ARE THERE ERRONEOUS RESULTS ON PATIENT SAMPLES FROM DIAGNOSTIC TEST? (IF YES OR UNKNOWN, GO TO QUESTION #2. IF NO, NO FURTHER QUESTION S REQUIRED.): NO.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD FACS¿ SAMPLE PREP ASSISTANT III CARRYOVER BETWEEN PATIENT SAMPLES HAD OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER THAT SPILLOVER SEEN IN SAMPLES. 1. ARE THERE ERRONEOUS RESULTS ON PATIENT SAMPLES FROM DIAGNOSTIC TEST? (IF YES OR UNKNOWN, GO TO QUESTION #2. IF NO, NO FURTHER QUESTION S REQUIRED.): NO

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2327509 BD FACSLYRIC¿ FLOW CYTOMETRIC REAGENTS AND ACCESSORIES OYE BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES 00382906628756

Patients

Seq Age Sex Outcome Treatment
1 Unknown