CARDIOCHEK PLUS ANALYZER
Report
- Report Number
- 1836135-2023-00027
- Event Type
- Malfunction
- Date Received
- July 28, 2023
- Date of Event
- June 30, 2023
- Report Date
- July 28, 2023
- Manufacturer
- PTS DIAGNOSTICS
- Product Code
- CGA
- UDI-DI
- 00381932701013
- PMA / PMN Number
- K140068
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER ALLEGATION OF OVERHEATING WAS NOT OBSERVED ON THE RETURNED ANALYZER. DURING THE INVESTIGATION, IT WAS DETERMINED THAT THE TERMINAL DISPLACEMENT IS MOST LIKELY CAUSED BY THE CUSTOMER PROVIDING UNDUE FORCE ONTO THE BATTERY TERMINALS. IN ADDITION, THE POTENTIAL EXISTS FOR THE BATTERY CONTACT TO LOSE TRACTION WITH THE COMPARTMENT CAUSING LESS FORCE TO DISLODGE HOWEVER IMPROPER REMOVAL OF BATTERIES WILL STILL FORCE THE DISLODGE. UPON VISUAL INSPECTION OF THE BATTERY COMPARTMENT, PLASTIC CASE DAMAGE AND PEN MARKINGS INDICATIVE OF IMPROPER/FORCEFUL REMOVAL OF THE BATTERIES WITH A FOREIGN OBJECT WAS OBSERVED. PTS IS CONFIDENT THAT THIS DAMAGE OCCURRED POST-DISTRIBUTION, AS THERE ARE NO MANUFACTURING STEPS THAT WOULD CAUSE THIS TYPE OF DAMAGE. NUMEROUS BATTERY COMPARTMENT IMPROVEMENTS HAVE ALREADY BEEN IMPLEMENTED PRIOR TO THESE ALLEGATIONS FROM THE FIELD. PROCESS IMPROVEMENTS AND PART SPECIFICATIONS HAVE ENHANCED SEATING AND HOLD CAPACITY OF THE BATTERY CONTACTS. ADDITIONALLY, A NEW BATTERY COMPARTMENT DESIGN (APPROVED BY THE FDA PER K193406) ELIMINATES THE POTENTIAL FOR OVERHEATING BY REMOVING THE POINT OF CONTACT THAT AN UNSEATED TERMINAL WOULD REACH.
THE CUSTOMER REPORTED THE ANALYZER WAS "WARM TO THE TOUCH". THE CUSTOMER REPORTED THE ANALYZER WOULD TURN OFF UPON FEELING WARM. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS. THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1967560 | CARDIOCHEK PLUS ANALYZER | CLINICAL CHEMISTRY ANALYZER | CGA | PTS DIAGNOSTICS | 00381932701013 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |