FDA Adverse Event Injury Summary report: N

COOLSCULPTING SYSTEM

MDR report key: 17410787 · Received July 27, 2023

Report

Report Number
3007215625-2023-11124
Event Type
Injury
Date Received
July 27, 2023
Report Date
November 14, 2023
Manufacturer
ALLERGAN (PLEASANTON)
Product Code
OOK
UDI-DI
00816417020001
PMA / PMN Number
K160259
Removal / Correction Number
Z-1347-2022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS IS A HISTORICAL RECORD AND REFLECTS REPORTS THAT WERE RECEIVED BY THE COMPANY PRIOR TO APRIL 2021. THIS MDR IS ASSOCIATED WITH A PARALLEL PLATE APPLICATOR, THESE APPLICATORS WERE DISCONTINUED AND RECALLED IN 2022. ACCORDING TO THE COOLSCULPTING USER MANUAL, UNDER RARE ADVERSE EVENTS, PARADOXICAL ADIPOSE HYPERPLASIA (PH/PAH) IS CHARACTERIZED BY A VISIBLY ENLARGED TISSUE VOLUME WITHIN THE TREATMENT AREA, WHICH MAY DEVELOP TWO TO FIVE MONTHS AFTER TREATMENT. SURGICAL INTERVENTION MAY BE REQUIRED. PAH IS NOT RELATED TO ANY COOLSCULPTING DEVICE FAILURE MODE BUT IT IS OBSERVED MORE FREQUENTLY WITH THE PARALLEL PLATE APPLICATORS. PAH IS INCLUDED IN THE RISK MANAGEMENT FILES OF THE DEVICE AS AN INHERENT RISK TO THE USE OF CRYOLIPOLYSIS FOR LOCALIZED FAT REDUCTION.

Description of Event or Problem · 0

ALLERGAN AESTHETICS RECEIVED A REPORT OF A PATIENT TREATED WITH COOLSCULPTING TO THE LOWER ABDOMEN ON (B)(6) 2015 USING A LEGACY COOLCORE APPLICATOR (BRZ-AP1-063-000) WHO DEVELOPED PARADOXICAL HYPERPLASIA (PAH/PH) OF THE ABDOMEN (ONSET NOT SPECIFIED) AND WAS DIAGNOSED ON (B)(6) 2016. (B)(6) AND (B)(6).

Description of Event or Problem · 0

ALLERGAN AESTHETICS RECEIVED A REPORT OF A PATIENT TREATED WITH COOLSCULPTING TO THE LOWER ABDOMEN ON (B)(6) 2015 USING A LEGACY COOLCORE APPLICATOR (BRZ-AP1-063-000) WHO DEVELOPED PARADOXICAL HYPERPLASIA (PAH / PH) OF THE ABDOMEN (ONSET NOT SPECIFIED) AND WAS DIAGNOSED ON (B)(6) 2016. (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1920697 COOLSCULPTING SYSTEM DERMAL COOLING PACK/VACUUM/MASSAGER OOK ALLERGAN (PLEASANTON) NI 00816417020001

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention (B)(6).