FDA Adverse Event Malfunction Summary report: N

CERTAIN¿ TITANIUM LARGE HEXED SCREW

MDR report key: 17408423 · Received July 27, 2023

Report

Report Number
0001038806-2023-01413
Event Type
Malfunction
Date Received
July 27, 2023
Date of Event
May 14, 2023
Report Date
January 29, 2024
Manufacturer
BIOMET 3I
Product Code
NHA
UDI-DI
00844868006138
PMA / PMN Number
K972444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). G4: PREMARKET IDENTIFICATION K972444/K072642.

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). ZIMVIE RECEIVED ONE (1) ILRGHT, (CERTAIN TITANIUM LARGE HEXED SCREW) FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED, A FRACTURE HAS BEEN IDENTIFIED AT THE THREADS. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1238798. NO DEVIATIONS OR NON-CONFORMANCE'S, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1238798 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : FUNCTIONAL : FRACTURE : SCREW BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WERE MATERIAL SELECTION OF THE PRODUCT IS NOT ADEQUATE TO WITHSTAND OCCLUSAL FORCES. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE FRACTURE HAS BEEN IDENTIFIED ON THE THREADS. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT 3 SCREWS AT TOOTH SITES 13, 14 AND 16 FRACTURED AND HAD TO BE REMOVED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1107109 CERTAIN¿ TITANIUM LARGE HEXED SCREW DENTAL SCREW NHA BIOMET 3I 1238798 00844868006138

Patients

Seq Age Sex Outcome Treatment
1 Male