BD SCHAEDLER AGAR WITH VITAMIN K1 AND 5% SHEEP BLOOD, PLATE, 90 MM X 120
Report
- Report Number
- 9680577-2023-00036
- Event Type
- Malfunction
- Date Received
- July 26, 2023
- Date of Event
- July 6, 2023
- Report Date
- July 24, 2023
- Manufacturer
- BECTON DICKINSON GMBH
- Product Code
- JSG
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
E.1. INITIAL REPORTER PHONE #: (B)(6). H.6. INVESTIGATION SUMMARY: THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR RECENT COMPLAINT (B)(4) AGAINST BD SCHAEDLER AGAR WITH VITAMIN K1 AND 5 % SHEEP BLOOD, CATALOG NUMBER 254084, LOT NUMBER 3122948 WITH RESPECT TO CONTAMINATION. EVENT DESCRIPTION: THE CUSTOMER REPORTED CONTAMINATION OF A PLATE. COMPLAINT HISTORY REVIEW: THE COMPLAINTS TRENDS WERE REVIEWED, AND NO SIMILAR COMPLAINT WAS REGISTERED FOR THIS LOT NUMBER. THUS, A TREND WAS NOT IDENTIFIED. BATCH HISTORY RECORD (BHR) REVIEW: THE BATCH HISTORY REVIEW DID NOT INDICATE ANY DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY, AND NO DEVIATIONS WERE OBSERVED. SAMPLE ANALYSIS: THE RETAIN SAMPLES WERE REVIEWED AND NO DEVIATION COULD BE DETECTED. RETURN SAMPLES WERE NOT PROVIDED; HOWEVER, PICTURES SAMPLES WERE SHARED DEMONSTRATING A CONTAMINATION. EVALUATION RESULTS: AT THIS STAGE OF OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMIC FAILURE IN OUR MANUFACTURING PROCESS. THIS PRODUCT DOES NOT HAVE AN SAL (STERILITY ASSURANCE LEVEL) CLAIM. IT IS FILLED ASEPTICALLY; THEREFORE, AN OCCURRENCE OF A CONTAMINATION EVENT CANNOT BE ENTIRELY RULED OUT. INVESTIGATION CONCLUSION: BASED ON THE EVALUATION AND THE PROVIDED PICTURES, THE COMPLAINT WAS CONFIRMED FOR CONTAMINATION. A DEFINITE ROOT CAUSE COULD NOT BE DETERMINED. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT WHILE USING BD SCHAEDLER AGAR WITH VITAMIN K1 AND 5% SHEEP BLOOD, PLATE, 90 MM X 120, ONE PLATE WAS CONTAMINATED. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CONTAMINATED PLATE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 662584 | BD SCHAEDLER AGAR WITH VITAMIN K1 AND 5% SHEEP BLOOD, PLATE, 90 MM X 120 | CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL | JSG | BECTON DICKINSON GMBH | 3122948 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |